Description
Product Overview
Parker ULC-512FF3RF-000 is a four channel radial Ultra Low Carryover Gradient Valve with 12 VDC coils, FFKM seals and a PEEK body. Its five female 1/4-28 ports place four normally closed channel paths around a central common port. The radial design has 9.4 µL seal-to-common-port volume, compared with 12.99 µL for the panel design, and uses two counterbored through holes for mounting.
Parker ULC-512FF3RF-000 Technical Specifications
- Manufacturer and family: Parker Hannifin, Ultra Low Carryover Gradient Valve
- Exact part number: ULC-512FF3RF-000
- Configuration: 4 Channel Radial, 4 normally closed channels and 5 liquid ports
- Wetted body and diaphragm: PEEK; FFKM
- Orifice: 0.030 inch (0.76 mm)
- Operating pressure: 45 psig (3.1 bar)
- Proof pressure: 120 psig (8.3 bar)
- Minimum water flow: 320 mL/min at 45 psig (3.1 bar), published Gradient family specification
- Air leakage: 0.150 sccm of air (bubble tight), published Gradient family specification
- Response: Less than 10 ms at rated voltage; less than 2 ms at twice rated voltage
- Coil electrical values: 12 VDC; 3.0 W nominal per coil; 250 mA published current per coil; 48 ohms ±10% at 20 °C
- Electrical interface: 8 UL1429-26-7 black irradiated PVC wire leads, 4.50 ±0.25 inch (114.3 ±6.35 mm); 4 latching Molex housings 50-57-9402 with contacts 16-02-0098
- Hit and hold: Hit and hold is required for actuation exceeding 100 ms. Use a rated-voltage hit longer than 20 ms and no longer than 100 ms, with 50 ms recommended, then 6 VDC hold at 0.70 W typical per coil
- Liquid interface: Four channel ports and one common port; 1/4-28 UNF-2B female threads
- Internal volume: 9.4 µL, seal to common port
- Published design weight: 2.56 oz (72.5 g)
- Reference dimensions: 49.5 × 49.5 × 30.16 mm drawing reference envelope, excluding leads and connectors
- Mounting: Radial body with two counterbored through holes
- Operating environment and media temperature: 4 to 50 °C
- Storage temperature: −20 to 70 °C
- Recommended filtration: 16 µm
- Reliability: 50 million cycles, published Gradient family rating
- Compliance and ingress rating: RoHS and REACH compliant; IP65, published Gradient family statements
Liquid Selection and Material Compatibility
4 normally closed liquid channels share one common port. A channel opens when its coil is energised and closes when that coil is de-energised. Coordinate channel timing with the pump and required solvent selection sequence. Parker rates the FFKM diaphragm material excellent with toluene and good with tetrahydrofuran and 12% bleach. PEEK is rated excellent with all three. These are material ratings; validate the complete valve with the actual liquid formulation, concentration, temperature and cleaning cycle.
Electrical and Mechanical Integration
Validate boosted actuation separately from the rated-voltage hit and hold profile.
Two 2.39 mm diameter through holes have 4.76 mm diameter counterbores with a depth of 2.54 mm. The optional 290-006061-005 wire extension provides 19.5 inch (495.3 mm) flying leads. Allow space for the liquid fittings and electrical connections.
Alternative Configurations
Compare these standard alternatives with the specifications above. Each entry highlights the changed attributes and gives the values for the named alternative.
- Parker ULC-524FF3RF-000: rated voltage: 24 VDC; published current per coil: 116 mA; nominal coil resistance: 207 ohms; hold voltage: 12 VDC; typical hold power per coil: 0.75 W.
- Parker ULC-512EP3RF-000: diaphragm and seal material: EPDM.
- Parker ULC-312FF34F-000: design: 2 Channel; channel count: 2; port count: 3; seal-to-common-port volume: 4.05 µL; published design weight: 33.7 g; reference envelope: 37.3 × 16.41 × 30.25 mm; mounting arrangement: Body support using the published mounting geometry; connector count: 2; lead wire count: 4.
- Parker ULC-512FF3SF-000: design: 4 Channel Panel; seal-to-common-port volume: 12.99 µL; published design weight: 69.5 g; reference envelope: 31.75 × 31.75 × 37.20 mm; mounting arrangement: Panel mount with two 4-40 UNC-2B threaded mounting holes.
OEM Analytical Fluid Control Support
Air Engineering Group Ltd can supply Parker ULC-512FF3RF-000 for HPLC solvent selection, HbA1c reagent handling and compatible analytical liquid circuits. Support includes valve selection, fluid circuit and manifold design, prototyping, assembly and functional testing against agreed requirements. Match the 4 channel radial design, 12 VDC control and FFKM seals to the complete instrument.
Due to the technical nature of this product, please contact our sales team on 01342 826488 or email info@aegroup.uk.com if you have any questions.
Brand
Parker
Parker Hannifin develops precision fluid control valves for OEM equipment that requires compact dimensions, reliable switching and efficient electrical control. Its extensive portfolio includes miniature pneumatic solenoid valves, proportional flow valves, cartridge valves and diaphragm isolation valves for controlling gases and liquids. Available configurations support pressure and vacuum switching, proportional flow control and media isolation, with options for manifold, PCB and tube integration.
Parker valve technologies are used within respiratory equipment, anaesthesia systems, patient monitoring devices, diagnostic instruments, laboratory automation and other medical or life science equipment. AEG helps OEM customers select the appropriate valve function, pressure range, flow capacity, voltage, materials, elastomer and connection method, before integrating the selected components into custom manifolds, pneumatic assemblies or complete fluid control solutions.
Lead Time
Product Availability
We work with a wide network of trusted manufacturers and suppliers to offer a comprehensive range of industrial components and systems. Many popular items are held in stock and available for prompt dispatch. These products are supplied directly from our UK facilities where possible to minimise lead times.
Order In and Made to Order Items
Some specialist products are not held as off the shelf stock. These items are ordered directly from the manufacturer and may be produced to order or configured for specific applications. As a result, lead times can be longer than standard stocked products, particularly for low volume or unique items. Lead times will vary depending on supplier schedules and production requirements.
Lead Times and Order Updates
For all order in products, we aim to provide clear lead time information at the point of order or shortly afterwards. Our team will keep you informed throughout the process and will advise if supplier lead times change. If you have a time critical requirement, we recommend contacting us before placing your order so we can confirm availability.
Delivery Charges
Additional delivery charges may apply to large orders, oversized items, palletised shipments, or deliveries outside the UK mainland including remote locations. Our sales team can confirm any applicable charges either before or after your order is placed to ensure full transparency.
Availability and Lead Time Enquiries
For the latest information on product availability and current lead times, please contact our team directly. We will confirm stock status, supplier lead times, and delivery expectations to support your ordering decision. You can email info@aegroup.uk.com or call +44 (0)1342 826488 to speak with a member of our team who can assist with your enquiry.
More Than Components
Experts you can trust. Technologies you can depend on.
Air Engineering Group is more than a component supplier. We operate as an engineering partner supporting customers from concept through to long term operation. Our role extends beyond product supply into system design, integration, build, installation, and lifecycle support. This approach allows us to solve complex technical challenges while improving reliability, efficiency, and operational assurance across demanding industrial environments.
Our OEM Fluid Technology Division focuses on engineered solutions for original equipment manufacturers. We support customers with design input, component selection, manifold assemblies, sub assemblies, and fully integrated pneumatic and fluid systems. Working closely with OEM teams, we help reduce complexity, improve performance, and accelerate development while maintaining consistency from prototype through to production.
The Compressed Air and Gas Division delivers complete compressed air and gas solutions for industrial and commercial applications. This includes system design, equipment supply, installation, commissioning, servicing, and ongoing maintenance. By taking responsibility for the full system, we help customers improve uptime, control energy use, and protect critical assets throughout their operational life.
Our Pneumatic and Fluid Product Division provides a comprehensive range of valves, actuators, regulators, filtration, fittings, and control components. Supported by in house technical expertise, this division ensures customers specify the correct products for their application while benefiting from reliable supply and informed engineering support. Together, these three divisions position us as a complete engineering solutions provider rather than a simple distributor.
Our approach to business is built around a clear customer first focus supported by engineering integrity and accountability. We take time to understand each application, operating environment, and performance requirement before recommending or designing a solution. This ensures customers receive systems and components that deliver reliability, efficiency, and long term value rather than short term fixes. Through clear communication, responsive support, and consistent technical standards, we build trusted partnerships that prioritise uptime, risk reduction, and operational confidence.
Speak to our engineering team by emailing info@aegroup.uk.com or call +44 (0)1342 826488 to speak with a member of our team who can assist with your enquiry.
Applications & Integrations
Applications and Integration
Parker ULC-512FF3RF-000 supports selection among four compatible HPLC solvents through the radial common-port layout. Its FFKM sealing and 9.4 µL seal-to-common-port volume suit evaluation of a four channel circuit where radial fitting access can be accommodated.
Parker identifies HPLC, HbA1c, reagent selection and distribution as applications for the Gradient family. The typical HPLC circuit places the valve in the solvent selection path upstream of the pump. Match the 45 psig rating to that liquid circuit.
Channel Control
Use 12 VDC rated coils. Hit and hold is required for actuation exceeding 100 ms. Use a rated-voltage hit longer than 20 ms and no longer than 100 ms, with 50 ms recommended, then 6 VDC hold at 0.70 W typical per coil. Drive the 4 channel coils separately and coordinate their normally closed operation with the pump sequence.
Installation and Liquid Compatibility
The radial body places the four channel ports around a central common port. Use the counterbored mounting holes and allow clearance around the coils, fittings and leads. Use 16 µm recommended filtration and keep both the media and environment within 4 to 50 °C. Validate PEEK and FFKM against the complete liquid formulation, concentration and cleaning cycle.