EXperts you can trust. Technologies you can depend on.

Experts you can trust. Technologies you can depend on.

Parker 018-0301-900 Series 18 4 Station PTFE Manifold Valve 9 VDC

Function: 4 stations; 2 way NC each
Pressure: 20 psig (1.38 bar); vacuum capable
Electrical: 9 VDC; 2.5 W published coil rating
Fluid ports: 5 × 1/4-28 UNF-2B
Wetted path and response: PTFE; <12 ms

This product is supplied to order, so pricing may vary. Contact us for up-to-date pricing and availability.

Not all items are held in stock as some are made to order or sourced on request. In-stock items usually arrive within 1–2 days, and we’ll confirm lead times by email once your order is processed.

Need help with this product or looking for an alternative? Talk to an expert today by emailing info@aegroup.uk.com or calling 01342 826488.

Description

Product Overview

Parker 018-0301-900 is a Series 18 miniature PTFE manifold valve with 4 independently controlled two-way normally closed stations and 9 VDC operation. It connects 4 liquid branches to one common port through an isolated PTFE wetted path. Maximum operating pressure is 20 psig, approximately 1.38 bar.

The four-station 9 VDC build is the only 9 VDC code in this ordering table. It supports selection or combination of four liquid branches through one common port. The published electrical table gives 273 mA and 33 ohms for this supply, distinguishing its driver requirements from the 12 VDC and 24 VDC four-station alternatives.

Alternative Configurations

  • Parker 018-0074-900: supply voltage changes from 9 VDC to 12 VDC; published nominal current changes from 273 mA to 211 mA; published nominal coil resistance changes from 33 ohms to 57 ohms.
  • Parker 018-0048-900: supply voltage changes from 9 VDC to 24 VDC; published coil input changes from 2.5 W to 4.2 W; published nominal current changes from 273 mA to 173 mA; published nominal coil resistance changes from 33 ohms to 139 ohms.

Parker 018-0301-900 Technical Specifications

  • Manufacturer and series: Parker Hannifin, Series 18
  • Exact part number: 018-0301-900
  • Valve configuration: 4 stations; two-way normally closed at each station
  • Fluid ports: 4 branch ports and one common port, 5 fluid ports in total
  • Connections: 1/4-28 UNF-2B female flat-bottom threaded ports
  • Operating pressure: Vacuum to 20 psig (approximately 1.38 bar)
  • Orifice: 0.062 inch (1.57 mm), published ordering value
  • Proof pressure: 1.5 times rated operating pressure
  • Wetted materials: PTFE body, seals and poppet diaphragms; all wetted parts PTFE
  • Electrical supply: 9 VDC
  • Published coil rating: 2.5 W; 273 mA nominal; 33 ohms ±5% at 21 °C
  • Electrical connection: 12 inch (304.8 mm) 26 AWG PTFE insulated lead wires
  • Response time: Less than 12 ms, published family specification
  • Leak performance: Bubble tight, published family specification
  • Operating environment: 4 to 66 °C; 100% continuous duty
  • Internal volume: Valve: 12.6 µL; inlet path: 25.2 µL; outlet path: 57.1 µL
  • Weight: 12.4 oz (352 g), published 4-station configuration figure
  • Life rating: 10 million cycles, application dependent
  • Recommended filtration: 10 µm maximum
  • Reference geometry: Drawing references: 104.4 mm span across opposed coils; 38.1 mm square central body; 31.8 mm body thickness; 20.6 mm coil diameter
  • Mounting: 2 mounting through holes, 4.1 mm diameter, in the central manifold body
  • Compliance: RoHS compliant, family statement

Fluid and Electrical Integration

Each of the 4 stations is independently controlled and normally closed. Its branch port is isolated from the common port when de-energised and opens to the common port when energised. Use the branch ports as inlets for flow mixing, or the common port as the inlet for selection and stream splitting.

Provide the rated 9 VDC supply and separate control for the four coils. Size the supply and driver for the number of stations energised together and assess coil heating in the instrument.

2 mounting through holes, 4.1 mm diameter, in the central manifold body. Use matching 1/4-28 UNF-2B fittings for the flat-bottom liquid ports. Allow space beyond the drawing references for fittings, connected tubing and lead routing. Select all valve and fitting materials for the liquid, concentration and temperature. Keep the manifold within its 20 psig rating, including when installed on the solvent side of an HPLC pump.

OEM Fluid Control Support

Air Engineering Group Ltd can supply Parker 018-0301-900 and support its 4-station layout and 9 VDC control within OEM liquid systems. Support includes valve selection, circuit and manifold design, prototyping, assembly and functional testing against agreed application requirements.

Due to the technical nature of this product, please contact our sales team on 01342 826488 or email info@aegroup.uk.com if you have any questions.

Brand

Parker

Parker Hannifin develops precision fluid control valves for OEM equipment that requires compact dimensions, reliable switching and efficient electrical control. Its extensive portfolio includes miniature pneumatic solenoid valves, proportional flow valves, cartridge valves and diaphragm isolation valves for controlling gases and liquids. Available configurations support pressure and vacuum switching, proportional flow control and media isolation, with options for manifold, PCB and tube integration.
Parker valve technologies are used within respiratory equipment, anaesthesia systems, patient monitoring devices, diagnostic instruments, laboratory automation and other medical or life science equipment. AEG helps OEM customers select the appropriate valve function, pressure range, flow capacity, voltage, materials, elastomer and connection method, before integrating the selected components into custom manifolds, pneumatic assemblies or complete fluid control solutions.

Technical Information

Technical Specifications

Series and ISO
Series 18

Voltage:
9 VDC

Valve Operation Type:
Electronic

Valve Actuation:
Solenoid actuated multi-station diaphragm isolation valve

Valve Function:
2 Way Normally Closed at each station

Operating Pressure:
20 psig

Orifice Size:
0.062 inch (1.57 mm), published ordering value

Primary Material:
PTFE

Filter Grade in micron:
10 µm maximum recommended

Manufacturer:
Parker Hannifin

Max Supply Pressure:
20 psig (approximately 1.38 bar)

Power Supply:
2.5 W, published coil rating

Response:
Less than 12 ms, published family specification

Temperature Range:
4 to 66 °C operating environment

Vacuum Range:
Vacuum capable

Operating Media:
Compatible liquids

ROHS Compliant:
YES
Temperature Max (Celsius):
66 °C
Temperature Minimum (Celsius) EU:
4 °C
Temperature Minimum (Fahrenheit):
66 °C

 

Air Engineering Group makes every effort to provide the most current product details directly from our manufacturers. However, because manufacturer specifications change frequently, we recommend verifying critical dimensions or performance metrics with our team prior to purchase.

 

Due to the technical nature of this product, please contact our sales team on 01342 826488 or email info@aegroup.uk.com if you have any questions.

Lead Time

Product Availability

We work with a wide network of trusted manufacturers and suppliers to offer a comprehensive range of industrial components and systems. Many popular items are held in stock and available for prompt dispatch. These products are supplied directly from our UK facilities where possible to minimise lead times.


Order In and Made to Order Items

Some specialist products are not held as off the shelf stock. These items are ordered directly from the manufacturer and may be produced to order or configured for specific applications. As a result, lead times can be longer than standard stocked products, particularly for low volume or unique items. Lead times will vary depending on supplier schedules and production requirements.


Lead Times and Order Updates

For all order in products, we aim to provide clear lead time information at the point of order or shortly afterwards. Our team will keep you informed throughout the process and will advise if supplier lead times change. If you have a time critical requirement, we recommend contacting us before placing your order so we can confirm availability.


Delivery Charges

Additional delivery charges may apply to large orders, oversized items, palletised shipments, or deliveries outside the UK mainland including remote locations. Our sales team can confirm any applicable charges either before or after your order is placed to ensure full transparency.


Availability and Lead Time Enquiries

For the latest information on product availability and current lead times, please contact our team directly. We will confirm stock status, supplier lead times, and delivery expectations to support your ordering decision. You can email info@aegroup.uk.com or call +44 (0)1342 826488 to speak with a member of our team who can assist with your enquiry.

More Than Components

Experts you can trust. Technologies you can depend on.

 

Air Engineering Group is more than a component supplier. We operate as an engineering partner supporting customers from concept through to long term operation. Our role extends beyond product supply into system design, integration, build, installation, and lifecycle support. This approach allows us to solve complex technical challenges while improving reliability, efficiency, and operational assurance across demanding industrial environments.

 

Our OEM Fluid Technology Division focuses on engineered solutions for original equipment manufacturers. We support customers with design input, component selection, manifold assemblies, sub assemblies, and fully integrated pneumatic and fluid systems. Working closely with OEM teams, we help reduce complexity, improve performance, and accelerate development while maintaining consistency from prototype through to production.

 

The Compressed Air and Gas Division delivers complete compressed air and gas solutions for industrial and commercial applications. This includes system design, equipment supply, installation, commissioning, servicing, and ongoing maintenance. By taking responsibility for the full system, we help customers improve uptime, control energy use, and protect critical assets throughout their operational life.

 

Our Pneumatic and Fluid Product Division provides a comprehensive range of valves, actuators, regulators, filtration, fittings, and control components. Supported by in house technical expertise, this division ensures customers specify the correct products for their application while benefiting from reliable supply and informed engineering support. Together, these three divisions position us as a complete engineering solutions provider rather than a simple distributor.

 

Our approach to business is built around a clear customer first focus supported by engineering integrity and accountability. We take time to understand each application, operating environment, and performance requirement before recommending or designing a solution. This ensures customers receive systems and components that deliver reliability, efficiency, and long term value rather than short term fixes. Through clear communication, responsive support, and consistent technical standards, we build trusted partnerships that prioritise uptime, risk reduction, and operational confidence.

 

Speak to our engineering team by emailing info@aegroup.uk.com or call +44 (0)1342 826488 to speak with a member of our team who can assist with your enquiry.

Applications & Integrations

Applications and Integration

Parker 018-0301-900 provides 4 normally closed liquid channels with 9 VDC control, 5 threaded fluid ports and a 20 psig pressure limit. The four-station 9 VDC build is the only 9 VDC code in this ordering table. It supports selection or combination of four liquid branches through one common port. The published electrical table gives 273 mA and 33 ohms for this supply, distinguishing its driver requirements from the 12 VDC and 24 VDC four-station alternatives.

Parker lists liquid chromatography gradient generation, liquid selection and mixing as family applications. Its symmetrical manifold design gives each fluid channel the same path length and timing characteristics.

Circuit and Mechanical Integration

Each of the 4 stations is independently controlled and normally closed. Its branch port is isolated from the common port when de-energised and opens to the common port when energised. Use the branch ports as inlets for flow mixing, or the common port as the inlet for selection and stream splitting.

The published volume breakdown is valve: 12.6 µl; inlet path: 25.2 µl; outlet path: 57.1 µl. Use these separate path values when reviewing delay volume. Provide the rated 9 VDC supply and separate control for the four coils. Size the supply and driver for the number of stations energised together and assess coil heating in the instrument.

Mount through the 2 holes of 4.1 mm diameter and allow access to all 5 liquid ports. Use matching 1/4-28 UNF-2B fittings for the flat-bottom liquid ports. Allow space beyond the drawing references for fittings, connected tubing and lead routing. Select all valve and fitting materials for the liquid, concentration and temperature. Use 10 µm maximum recommended filtration and keep the operating environment within 4 to 66 °C. The manifold’s working limit remains 20 psig when used upstream of an HPLC pump.

Explore Our Related Parker OEM Capabilities:

Parker Precision Miniature Valve Solutions
Learn how AEG supplies and integrates Parker miniature solenoid and proportional valves.
Follow the process from application review and concept development through to prototyping, manufacture and assembly.
OEM Pneumatic and Fluid Control Solutions
Explore our capabilities in component integration, custom manifolds and complete OEM assemblies.
Medical Device OEM Solutions
Discover our support for medical device fluid control systems, subassemblies and repeat production.

Associated Products

Components Integrated into OEM Solutions

Our capabilities extend beyond individual component supply. We integrate pneumatic and fluid control products into custom manifolds, control assemblies and complete OEM solutions developed around the specific technical, operational and production requirements of each machine, device or system.

AEG can assist from initial concept and component selection through to prototyping, assembly and testing. By combining proven products with practical design expertise, we help OEMs reduce connections, simplify installation, improve repeatability and prepare reliable solutions for ongoing production.