Product Overview
Parker 001-0028-900 is a Series 1 miniature PTFE diaphragm isolation valve with three-way switching, 24 VDC control and 4.2 W rated power. It has a 0.060 inch (1.52 mm) orifice, 1/4-28 UNF-2B female fluid ports and a 20 psig (approximately 1.38 bar) pressure limit. The wetted liquid path is isolated from the solenoid.
This three-way build provides common-line switching for instruments with a 24 VDC supply. It retains the 96 µL internal volume and less than 12 ms cycling response of the 12 VDC Series 1 option. Nominal input increases to 4.2 W; after actuation, the published 12 VDC hold setting uses 1.04 W.
Alternative Configurations
- Parker 001-0017-900: supply voltage changes from 24 VDC to 12 VDC; rated power changes from 4.2 W to 2.5 W; nominal current changes from 173 mA to 211 mA; nominal coil resistance changes from 139 ohms to 57 ohms; hold voltage changes from 12 VDC to 6 VDC; hold power changes from 1.04 W to 0.63 W.
- Parker 002-0010-900: series changes from Series 1 to Series 2; function changes from 3 Way to 2 Way NC; wetted path changes from PTFE body and diaphragm; borosilicate glass to PTFE body and diaphragm; internal volume changes from 96 µL to 49 µL; cycling response changes from less than 12 ms cycling to less than 20 ms cycling; hold voltage changes from 12 VDC to 8 VDC; hold power changes from 1.04 W to 0.46 W.
Parker 001-0028-900 Technical Specifications
- Manufacturer and series: Parker Hannifin, Series 1
- Exact part number: 001-0028-900
- Valve type and function: Direct acting diaphragm isolation valve; 3 Way
- Operating pressure: Vacuum to 20 psig (approximately 1.38 bar)
- Proof pressure: 1.5 times rated operating pressure
- Orifice: 0.060 inch (1.52 mm)
- Fluid connections: Three 1/4-28 UNF-2B female ports
- Wetted materials: PTFE body and diaphragm; borosilicate glass
- Electrical rating: 24 VDC, 4.2 W, 173 mA nominal, 139 ohms ±5% at 21.1 °C
- Electrical connection: 26 AWG PTFE insulated lead wires
- Cycling response: Less than 12 ms cycling
- Leak performance: Bubble tight, published family specification
- Operating environment: 4 to 66 °C; 100% continuous duty
- Internal volume: 96 µL, published 3 way specification
- Life rating: 10 million cycles, application dependent
- Weight: 2.7 oz (76.5 g), published family figure
- Recommended filtration: 10 µm min, Parker recommendation
- Reference dimensions: 51.3 mm drawing reference height; 25.4 mm body diameter; 20.6 mm coil diameter
- Mounting: Four #6-32 UNC-2B mounting holes equally spaced on a 17.78 mm diameter bolt circle
- Compliance: RoHS compliant, family statement
Fluid and Electrical Integration
When de-energised, common port 2 connects to normally open port 3 and port 1 is closed. Energising the coil connects common port 2 to normally closed port 1 and closes port 3. Direct acting operation requires no applied pressure or vacuum to actuate the valve.
Apply the full rated 24 VDC for initial actuation, then reduce to the published 12 VDC hold setting at 1.04 W. Validate switching and holding timing in the assembled circuit. Support the insulated leads and use matching 1/4-28 fittings suitable for the liquid circuit.
Four #6-32 UNC-2B mounting holes equally spaced on a 17.78 mm diameter bolt circle. Allow the illustrated 6.4 mm clearance above the coil and additional space for lead routing, fittings and connected tubing. Check both PTFE and borosilicate glass against the liquid, concentration and temperature. Parker rates borosilicate glass as not recommended for toluene and 20% sodium hydroxide in its abbreviated material chart.
OEM Fluid Control Support
Air Engineering Group Ltd can supply Parker 001-0028-900 and support its 3 way function and 24 VDC control within OEM fluid equipment. Support includes valve selection, fluid circuit and manifold design, prototyping, assembly and functional testing against agreed application requirements.
Due to the technical nature of this product, please contact our sales team on 01342 826488 or email info@aegroup.uk.com if you have any questions.
Brand
Parker
Parker Hannifin develops precision fluid control valves for OEM equipment that requires compact dimensions, reliable switching and efficient electrical control. Its extensive portfolio includes miniature pneumatic solenoid valves, proportional flow valves, cartridge valves and diaphragm isolation valves for controlling gases and liquids. Available configurations support pressure and vacuum switching, proportional flow control and media isolation, with options for manifold, PCB and tube integration.
Parker valve technologies are used within respiratory equipment, anaesthesia systems, patient monitoring devices, diagnostic instruments, laboratory automation and other medical or life science equipment. AEG helps OEM customers select the appropriate valve function, pressure range, flow capacity, voltage, materials, elastomer and connection method, before integrating the selected components into custom manifolds, pneumatic assemblies or complete fluid control solutions.
Lead Time
Product Availability
We work with a wide network of trusted manufacturers and suppliers to offer a comprehensive range of industrial components and systems. Many popular items are held in stock and available for prompt dispatch. These products are supplied directly from our UK facilities where possible to minimise lead times.
Order In and Made to Order Items
Some specialist products are not held as off the shelf stock. These items are ordered directly from the manufacturer and may be produced to order or configured for specific applications. As a result, lead times can be longer than standard stocked products, particularly for low volume or unique items. Lead times will vary depending on supplier schedules and production requirements.
Lead Times and Order Updates
For all order in products, we aim to provide clear lead time information at the point of order or shortly afterwards. Our team will keep you informed throughout the process and will advise if supplier lead times change. If you have a time critical requirement, we recommend contacting us before placing your order so we can confirm availability.
Delivery Charges
Additional delivery charges may apply to large orders, oversized items, palletised shipments, or deliveries outside the UK mainland including remote locations. Our sales team can confirm any applicable charges either before or after your order is placed to ensure full transparency.
Availability and Lead Time Enquiries
For the latest information on product availability and current lead times, please contact our team directly. We will confirm stock status, supplier lead times, and delivery expectations to support your ordering decision. You can email info@aegroup.uk.com or call +44 (0)1342 826488 to speak with a member of our team who can assist with your enquiry.
More Than Components
Experts you can trust. Technologies you can depend on.
Air Engineering Group is more than a component supplier. We operate as an engineering partner supporting customers from concept through to long term operation. Our role extends beyond product supply into system design, integration, build, installation, and lifecycle support. This approach allows us to solve complex technical challenges while improving reliability, efficiency, and operational assurance across demanding industrial environments.
Our OEM Fluid Technology Division focuses on engineered solutions for original equipment manufacturers. We support customers with design input, component selection, manifold assemblies, sub assemblies, and fully integrated pneumatic and fluid systems. Working closely with OEM teams, we help reduce complexity, improve performance, and accelerate development while maintaining consistency from prototype through to production.
The Compressed Air and Gas Division delivers complete compressed air and gas solutions for industrial and commercial applications. This includes system design, equipment supply, installation, commissioning, servicing, and ongoing maintenance. By taking responsibility for the full system, we help customers improve uptime, control energy use, and protect critical assets throughout their operational life.
Our Pneumatic and Fluid Product Division provides a comprehensive range of valves, actuators, regulators, filtration, fittings, and control components. Supported by in house technical expertise, this division ensures customers specify the correct products for their application while benefiting from reliable supply and informed engineering support. Together, these three divisions position us as a complete engineering solutions provider rather than a simple distributor.
Our approach to business is built around a clear customer first focus supported by engineering integrity and accountability. We take time to understand each application, operating environment, and performance requirement before recommending or designing a solution. This ensures customers receive systems and components that deliver reliability, efficiency, and long term value rather than short term fixes. Through clear communication, responsive support, and consistent technical standards, we build trusted partnerships that prioritise uptime, risk reduction, and operational confidence.
Speak to our engineering team by emailing info@aegroup.uk.com or call +44 (0)1342 826488 to speak with a member of our team who can assist with your enquiry.
Applications & Integrations
Applications and Integration
Parker 001-0028-900 combines 3 way operation, 24 VDC control, a 0.060 inch orifice and a 20 psig maximum operating pressure. This three-way build provides common-line switching for instruments with a 24 VDC supply. It retains the 96 µL internal volume and less than 12 ms cycling response of the 12 VDC Series 1 option. Nominal input increases to 4.2 W; after actuation, the published 12 VDC hold setting uses 1.04 W.
Parker lists reagent control, solvent management and aggressive liquid control as applications for this valve family. The isolated fluid path separates the liquid from the solenoid.
Circuit Integration
When de-energised, common port 2 connects to normally open port 3 and port 1 is closed. Energising the coil connects common port 2 to normally closed port 1 and closes port 3. The 96 µL internal volume is the published value for this 3 way configuration. Check both PTFE and borosilicate glass against the liquid, concentration and temperature. Parker rates borosilicate glass as not recommended for toluene and 20% sodium hydroxide in its abbreviated material chart.
Provide 24 VDC and capacity for the rated 4.2 W input. Apply the full rated 24 VDC for initial actuation, then reduce to the published 12 VDC hold setting at 1.04 W. Validate switching and holding timing in the assembled circuit. Use the 26 AWG PTFE insulated lead wires with suitable support.
Use matching 1/4-28 UNF-2B fluid fittings and the four #6-32 UNC-2B body mounting holes on a 17.78 mm diameter bolt circle. Allow the illustrated 6.4 mm clearance above the coil and additional space for lead routing, fittings and connected tubing. Keep the operating environment within 4 to 66 °C and follow Parker’s 10 µm min filtration recommendation.