Parker SRS Miniature Pneumatic Solenoid Valves for OEMs
Compact manifold mounted gas switching for medical devices and portable instruments
Air Engineering Group supplies Parker SRS miniature pneumatic solenoid valves for OEM gas control assemblies. The range uses a three port, direct acting poppet design with normally closed, normally open, and distributor functions, allowing supply, exhaust, or routing control within a manifold.
Choose the pressure model and valve function together. The published models have different orifices, power ratings, vacuum limits, and life references. Assess the selected configuration against inlet and outlet pressures, required flow, electrical supply, and the intended operating sequence.
The SRS combines a compact body with an integrated manifold seal arrangement and electrical connection options. Plan the manifold, wiring, and service access around the selected version, then qualify the assembled device for leakage, switching, and thermal behaviour.
How the SRS valve operates
Electrical energisation changes the poppet position and the connection between the common port and the other two ports. For pressure service, normally closed and normally open describe the supply path when power is removed. Distributor operation routes a common supply between two circuit paths.
Parker provides separate pressure and vacuum diagrams. Use the selected function’s porting information to define both electrical states and the required behaviour during power loss. Confirm the complete switching sequence with the actual connected volume and operating conditions.
What manifold mounting provides
The valve connects directly to passages in a manifold through the supplied gasket. This arrangement can shorten external tubing runs and bring switching closer to the controlled pressure or measurement point. Include the manifold passages when assessing total connected volume and response time.
Design port positions, sealing surfaces, fastening, electrical clearance, and access for replacement before releasing the assembly. Verify gasket placement, port isolation, and leakage after mounting. Include filters and any shared manifold passages in circuit qualification.
Parker SRS Miniature Pneumatic Solenoid valve OEM applications
Blood pressure monitoring
Assess inflation, isolation, and venting against cuff volume, tubing resistance, and the required pressure profile. Select the function for each circuit state, then qualify leakage and pressure response. Include power interruption, repeated switching, and internal device temperature.
Sensor zeroing
Define the reference pressure or gas path required during zeroing. Confirm the valve’s port assignments, leakage allowance, and settling time at the sensor. Test the zeroing sequence with the final manifold and tubing, including the effects of residual pressure and connected volume.
Patient monitoring equipment
Assess the valve’s role within the device’s gas or pressure circuit, including switching timing, contamination protection, and the required power loss state. Qualify performance with the intended filters, connections, and supply variation. Include simultaneous operation where several channels share a supply or manifold.
Portable analytical instruments
Compare valve power and switching duty with the instrument’s battery capacity and thermal budget. For gas sampling paths, assess leakage, residual gas, and sample arrival time. Validate switching and purge sequences at the measurement point with the intended gas and complete circuit.
Parker SRS Pneumatic Solenoid Valve Benefits
Compact manifold integration
The narrow body provides an option for closely grouped pneumatic channels. Assess the complete manifold footprint alongside fastening, wiring, filter access, and service clearance. The finished assembly dimensions depend on these connections as well as the valve body.
Choice of circuit function
Normally closed, normally open, and distributor arrangements provide different pressure circuit states. Select using the required behaviour when energised and de-energised, then confirm the port connections for pressure or vacuum service.
Electrical connection options
Pins and insulated leads are available in front or manifold interface positions. Choose the termination to match the PCB, harness, connector access, and assembly method. Include strain relief and clearance in the device layout.
Model selection for power and duty
The performance table links orifice, pressure, vacuum, power, and life requirements. Use the complete model specification when comparing options, then validate the selected valve under the intended gas exposure, electrical drive, and operating duty.
Parker SRS Valve Technical specifications
Published specifications from Parker’s SRS technical literature.
| Parameter | Published specification |
|---|---|
| Valve design | Three port, direct acting poppet |
| Functions | Normally closed, normally open, or distributor |
| Mounting | Manifold mount; gasket supplied |
| Basic dimensions | 38.1 mm length, 10.0 mm width, 15.5 mm height |
| Published weight | 6.5 g |
| Body and bobbin | Glass reinforced PBT |
| Pole and plunger | 430 FR stainless steel |
| Seal | FKM |
| Other wetted metal | 300 Series stainless steel |
| Media | Non-reactive gases |
| Voltage options | 5, 12, or 24 VDC, ±10% |
| Power options | 0.5 or 1.0 W, according to model |
| Published air leakage | Less than 0.016 sccm |
| Cycling response | Less than 30 ms |
| Operating environment | 0°C to 55°C |
| Storage temperature | −40°C to 70°C |
| Published internal volume | 0.027 cm³ |
| Recommended filtration | 40 µm |
| Published burst reference | 200 psig, 13.7 bar |
Allow additional space for the selected electrical connection, mating parts, and service access. Confirm the installation envelope against the current drawing. Use the leakage and response figures as qualification inputs, then validate performance within the device.
Model performance parameters
Confirm the inlet, outlet, and differential pressure limits for the intended operating conditions. Use the model’s operating limit; the burst reference does not establish a permitted service pressure.
The vacuum figures are vacuum readings, not absolute pressure values. Model 16 has a lower published vacuum limit than the other listed models. Treat the life requirements as model specific qualification references and validate service life with the actual circuit and duty.
| Model | Nominal orifice | Maximum supply pressure | Maximum supply vacuum | Power | Life requirement, million cycles |
|---|---|---|---|---|---|
| 10 | 0.51 mm | 35 psi / 2.41 bar | 27 inHg / 686 mmHg | 0.5 W | 175 |
| 11 | 0.51 mm | 85 psi / 5.86 bar | 27 inHg / 686 mmHg | 1.0 W | 50 |
| 13 | 0.76 mm | 20 psi / 1.37 bar | 27 inHg / 686 mmHg | 0.5 W | 200 |
| 14 | 0.76 mm | 50 psi / 3.44 bar | 27 inHg / 686 mmHg | 1.0 W | 25 |
| 16 | 1.14 mm | 10 psi / 0.68 bar | 20 inHg / 508 mmHg | 0.5 W | 100 |
| 17 | 1.14 mm | 20 psi / 1.37 bar | 27 inHg / 686 mmHg | 1.0 W | 25 |
These figures reproduce Parker’s model table, including its rounded metric pressure values.
The above component specifications come from Parker’s SRS Miniature Solenoid Valve datasheet.
Learn more about Parker Miniature Valves and Components
Parker SRS Flow selection
Parker’s typical flow curves use air at 24°C. Evaluate the curve for the selected orifice and model limit, accounting for inlet and outlet pressures. State the gas and reference conditions when specifying flow in standard litres per minute.
Measure circuit performance with the intended filter, manifold, tubing, and fittings. The published internal valve volume excludes these connected passages, which affect pressure rise, evacuation, and gas arrival time. Verify response at the device’s controlled or measurement point.
Materials and contamination control
Review the glass reinforced PBT, FKM, stainless steel, manifold materials, and gasket against the gas and anticipated contamination. Include humidity, condensate, cleaning residues, temperatures, and storage conditions when assessing compatibility.
Parker describes a hermetically sealed coil to help protect against accidental liquid exposure. Assess environmental protection for the complete assembly, including the connections and manifold. Prevent liquids entering the pneumatic path of a valve intended for non-reactive gas service.
Parker SRS Manifold and electrical integration
Manifold seal and mounting
Follow Parker’s current mounting drawing for port positions, seal contact surfaces, and fastening. Check gasket condition and alignment before assembly, and include the specified screws and any required test manifold in the bill of materials.
Control machining, deburring, and cleanliness to protect the seal and avoid cross-port leakage. Verify mounting loads and assembly clearances, then leak test the installed valve under the intended pressure and vacuum conditions. Provide access for inspection, filter changes, and valve replacement.
Electrical interfaces and reduced power control
Parker lists front or manifold interface versions of 0.025 inch square pins and insulated wire leads. Select the connection alongside supply voltage, coil power, harness routing, and PCB clearance. Confirm lead dimensions and termination details against the selected drawing.
Spike and hold or PWM control may reduce electrical demand. The SRS datasheet does not specify numerical hit time, hold voltage, PWM frequency, or duty cycle. Obtain guidance for the selected model, then validate opening, holding margin, release timing, and enclosure temperature with the proposed driver.
Prototype qualification and production selection
the right configuration for procurement
Qualify flow, leakage, switching, pressure response, and temperature with the intended circuit. Test simultaneous operation and power loss under representative conditions. Include sensor settling, venting, or vacuum recovery targets where required.
Specify the Parker part number, pressure model, valve function, voltage, and electrical interface. Include gas composition, pressures, flow, temperatures, filtration, connected volume, and duty.
Confirm the drawing, porting, gasket, and accessories. Provide development quantities and qualification timescales, plus production forecasts, delivery requirements, and documentation. Assess assembly and lifecycle costs, and review substitutions or changes with engineering.
Discuss your OEM gas control application
AEG’s OEM Fluid Technology Division can guide component selection and assembly development, considering gas compatibility, pressure, vacuum, flow, valve function, electrical drive, manifold sealing, filtration, and maintenance requirements.
Share your schematic, gas composition, inlet and outlet pressures, flow, switching sequence, electrical supply, temperatures, connected volume, and quantities. Include sensor settling, venting, or vacuum targets so we can assess SRS configurations and integration options.
Browse the Parker SRS pneumatic valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.
Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.
From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.
Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.