Parker Series 99 Dispense Valves for OEM Instruments
Miniature high speed liquid valves for pressurised dispensing, refrigerant control, and OEM fluid circuits
As a supplier of Parker precision fluid control products, Air Engineering Group brings practical experience in specifying miniature valves for medical and analytical OEM fluid control assemblies.
Parker Series 99 is a miniature solenoid valve for fast liquid shutoff, switching, and dispensing. The range includes two way normally closed valves and three way configurations for selecting between fluid routes. Parker identifies liquid CO₂, surgical refrigerant, and semiconductor refrigerant dispensing as typical applications.
Series 99 uses a non-isolation mechanism with a 316 stainless steel body and configuration dependent wetted materials. Published maximum pressure reaches 1250 psig, approximately 86.2 bar, for the 0.030 inch two way version. Larger orifices and different switching functions have lower limits, so specify the exact configuration rather than the family maximum.
How Parker Series 99 dispense valves work
The electromagnetic coil moves the internal armature and sealing mechanism to change the liquid flow path. A two way normally closed valve opens when energised and closes when power is removed. A three way valve switches its common connection between the normally open and normally closed routes.
A timed electrical command controls when the path changes. Delivered liquid volume still depends on pressure differential, viscosity, electrical drive, connecting volume, and the outlet restriction. For refrigerants, pressure reduction can also cause phase change and cooling, so validate the resulting delivery at the intended outlet.
| Electrical state | Two way normally closed | Three way |
|---|---|---|
| Coil deenergised | Flow path closed. | Common port 2 connects to normally open port 3; port 1 is closed. |
| Coil energised | Flow path open. | Common port 2 connects to normally closed port 1; port 3 is closed. |
Assign connections using Parker’s port diagram and the selected part drawing. Assess both the normal operating sequence and the fluid path maintained when electrical power is removed.
Medical, analytical, and life science applications
Medical and surgical refrigerant systems
Parker identifies surgical refrigerant dispensing as a typical application. OEM designers should assess the refrigerant, inlet pressure, discharge conditions, and switching sequence together. Qualify temperature behaviour, leakage, and delivery performance within the intended medical device and its representative operating and cleaning conditions.
Analytical liquid control instruments
Series 99 is designed for liquid analysis and can be evaluated for fast shutoff or route selection in pressurised analytical circuits. Define inlet and outlet pressures, liquid properties, and delivery timing, then validate the selected valve with the instrument’s connected tubing and dispensing geometry.
Life science and laboratory OEM equipment
For life science and laboratory equipment, assess Series 99 where the liquid chemistry and pressure conditions justify a stainless steel wetted valve. Suitability depends on every wetted material, the required switching function, and representative testing of cleaning, residual liquid, and delivery performance within the instrument.
Semiconductor and refrigerant dispensing
Parker also identifies liquid CO₂ and semiconductor refrigerant dispensing. Consider the valve alongside the supply, downstream restriction, and phase change behaviour. Specify the selected configuration against its pressure limit and verify delivery performance across the intended temperature, discharge conditions, and equipment operating sequence.
Why specify Parker Series 99 dispense valves?
Pressure capability matched to the configuration
The available orifices provide different combinations of flow and operating pressure. Selecting the largest passage is not automatically the best choice for a pressurised liquid circuit. Confirm the limit for the exact valve function and part number, then qualify flow and delivery under representative operating conditions.
Fast actuation for timed liquid switching
Parker publishes response times below 5 ms for the 0.030 and 0.060 inch orifices, and below 6 ms for the 0.116 inch version. These values inform switching design. Delivered volume still depends on pressure, liquid properties, electrical drive, and the connected fluidic circuit during instrument operation.
Documented wetted materials for selection
Passivated stainless steel construction provides a documented basis for chemical compatibility assessment. The wetted mechanism also includes other materials, with sealing arrangements dependent on the configuration. Review the full material list against process liquids and cleaning agents before specifying the valve for an OEM instrument’s operating conditions.
Defined leakage and thermal specifications
Parker publishes helium leakage testing and continuous duty operation across the stated ambient temperature range. These specifications help designers establish qualification criteria. Confirm how the selected configuration behaves within the assembly, including coil heating, liquid exposure, and the instrument’s required shutoff performance under representative operating conditions.
Match orifice and valve function to operating pressure
Parker’s ordering table distinguishes the pressure limits of two way and three way configurations. Use the exact valve function as well as orifice size when selecting a part.
| Orifice diameter | Valve function | Maximum operating pressure | Published response |
|---|---|---|---|
| 0.030 inch / 0.76 mm | Two way normally closed | 1250 psig / 86.2 bar | Less than 5 ms |
| 0.060 inch / 1.52 mm | Two way normally closed | 250 psig / 17.2 bar | Less than 5 ms |
| 0.060 inch / 1.52 mm | Three way | 100 psig / approximately 6.9 bar | Less than 5 ms |
| 0.116 inch / 2.95 mm | Two way normally closed | 100 psig / approximately 6.9 bar | Less than 6 ms |
| 0.116 inch / 2.95 mm | Three way | 100 psig / approximately 6.9 bar | Less than 6 ms |
Parker lists a lower vacuum limit of 1 × 10⁻⁵ Torr. Confirm the chosen part’s pressure and vacuum suitability for the intended duty. Proof pressure is 1.5 times rated pressure and is not an operating allowance.
Series 99 liquid valve specifications
The series offers documented operating limits and configuration choices for demanding liquid control. These give designers a basis for development testing and help procurement teams specify the required component accurately.
The following values are published in Parker’s Series 99 liquid datasheet. Confirm the selected part’s current specification and drawing before finalising an OEM design.
| Specification | Published value |
|---|---|
| Valve configurations | Two way normally closed or three way |
| Body material | 316 stainless steel |
| Primary sealing materials | Vespel for the 0.030 inch version; FKM for the 0.060 and 0.116 inch versions, with metallic gaskets |
| Other wetted materials | PTFE and stainless steel; confirm the full selected part material list |
| Standard coil voltages | 12 or 24 VDC; availability depends on configuration |
| Published rated coil power | 12 W at either standard voltage |
| Published coil current | 1 A at 12 VDC; 0.5 A at 24 VDC |
| Ambient operating temperature | 4 to 105°C |
| Duty rating | 100% continuous duty |
| Internal volume | 354.5 to 593.8 µL; configuration details require confirmation |
| Published helium leakage specification | 1 × 10⁻⁸ cc/sec/atm |
| Recommended filtration | 40 µm maximum |
| Porting | A-LOK compression fittings or female VacuSeal, depending on configuration |
The helium leakage figure is a published gas test specification. Establish liquid shutoff, delivery, and residual leakage acceptance criteria for the actual instrument separately.
The above component specifications come from Parker’s Series 99 Liquid Valve datasheet.
Learn more about Parker Miniature Valves and Components
Integrate Series 99 into an OEM dispensing circuit
Qualify pressure and the entire fluid circuit
Check the selected valve function, orifice, materials, and pressure rating alongside tubing, fittings, reservoirs, and downstream restrictions. Use representative inlet and outlet pressures for flow and delivery testing. Parker’s typical flow curves were measured with water at 24°C.
For refrigerants, assess phase change and cooling through the discharge path. Include filtration, leakage, flushing, and residual liquid in qualification. Validate delivery at the intended outlet, since connecting volume and nozzle geometry can affect the result.
Match electrical drive to delivery timing
Assess supply capacity and enclosure temperature against the published 12 W coil rating and the number of valves energised together. Parker documents a hit and hold method using 25 ms at rated voltage, followed by the holding settings below.
Validate opening, release, and delivery timing with the actual driver and coil suppression. The holding requirements differ between two way and three way configurations. Check the resulting liquid delivery under representative pressure, temperature, and operating sequences during prototype qualification.
| Valve function and rated coil | Hold voltage | Published hold power |
|---|---|---|
| Two way, 12 VDC | 5 V | 2.1 W |
| Two way, 24 VDC | 5 V | 0.52 W |
| Three way, 12 VDC | 6 V | 3 W |
| Three way, 24 VDC | 12 V | 3 W |
The published response time describes valve switching. It does not establish a minimum dispensing pulse, delivered volume, or dosing accuracy.
Parker's Published standard part numbers
Parker’s liquid datasheet lists the following configurations. The 0.030 inch version uses 1/16 inch A-LOK connections, the 0.060 inch version uses 1/8 inch A-LOK connections, and the 0.116 inch version uses 1/4 inch female VacuSeal connections. Confirm current availability when ordering.
| Valve function | Orifice diameter | Coil voltage | Parker part number |
|---|---|---|---|
| Two way normally closed | 0.030 inch / 0.76 mm | 12 VDC | 099-0051-900 |
| Two way normally closed | 0.030 inch / 0.76 mm | 24 VDC | 099-0340-900 |
| Two way normally closed | 0.060 inch / 1.52 mm | 24 VDC | 099-0080-900 |
| Three way | 0.060 inch / 1.52 mm | 12 VDC | 099-0075-900 |
| Three way | 0.060 inch / 1.52 mm | 24 VDC | 099-0135-900 |
| Two way normally closed | 0.116 inch / 2.95 mm | 24 VDC | 099-0167-900 |
| Three way | 0.116 inch / 2.95 mm | 24 VDC | 099-0107-900 |
These are published configuration references, with current availability requiring confirmation.
Specify with confidence from prototype to production
the right configuration for procurement
Specify the exact Parker part number alongside switching function, orifice, wetted materials, coil voltage, porting, and mounting requirements. Confirm pressure rating, electrical connection, and installation space against the selected part’s current drawing, rather than relying on the family’s maximum headline values.
For development enquiries, provide sample quantities, process and cleaning media, inlet and outlet pressures, flow requirements, delivery targets, and the qualification timetable. For production sourcing, include forecast volumes, delivery requirements, documentation, and agreed restrictions on substitutions.
Assess acquisition cost alongside driver hardware, pressure rated fittings, assembly time, qualification, and maintenance. Confirm lead times and current availability for the selected configuration.
Discuss your OEM dispensing valve requirements
AEG’s OEM Fluid Technology Division combines practical experience in medical and analytical fluid control with knowledge of miniature valve specification, material compatibility, and system integration. Our team can guide component selection and assembly development, considering switching function, pressure, orifice size, wetted materials, electrical drive, and delivery performance.
Share your fluidic schematic, process and cleaning media, inlet and outlet pressures, flow or dispense volume, pulse duration and frequency, electrical supply, temperatures, and quantities so we can assess your requirements and recommend options.
Browse the Parker Series 99 valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.
Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.
From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.
Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.