Parker Series 9 Calibrant Valves for OEM Instruments
Miniature gas and liquid switching for analytical, medical, and life science equipment
As a supplier of Parker precision fluid control products, Air Engineering Group helps OEMs specify miniature valves for medical and analytical equipment, considering both the component and its role within the assembled fluid circuit.
The Parker Series 9 is a miniature solenoid valve family for opening, closing, or selecting gas and liquid paths. Parker identifies calibrant gas control for mass spectrometers as a typical application. A calibrant is a reference material used to check or adjust an instrument’s measurement response.
Series 9 includes two way normally closed and three way configurations, with a 316 stainless steel body and configuration dependent sealing materials. Its published gas specifications combine fast switching, low helium leakage, and pressure ratings up to 1250 psig for the 0.030 inch two way version.
OEM selection should match the required switching function, media, pressure differential, leakage target, and electrical drive. Parker publishes separate gas and liquid datasheets, so use the information relevant to the selected duty.
What a Series 9 calibrant valve does
A calibrant valve controls the connection between a reference source and an analytical flow path. It can admit reference gas during a calibration sequence or select between connected sources, depending on the valve configuration.
Series 9 uses an electromagnetic coil to move a wetted internal armature and sealing mechanism. Parker describes it as a non-isolation valve. Compatibility therefore depends on the full wetted material arrangement, including the seals and moving components.
How the switching function works
In a two way normally closed valve, the flow path is closed without electrical power and opens when the coil is energised. A three way valve changes the connection between its common port and the normally open or normally closed port.
| Configuration | Coil deenergised | Coil energised |
|---|---|---|
| Two way normally closed | Flow path closed | Flow path open |
| Three way | Common port 2 connected to normally open port 3 | Common port 2 connected to normally closed port 1 |
The delivered calibrant conditions depend on source pressure, restrictions, gas properties, connecting volume, and the instrument’s response. Where a vapour source is used, temperature and condensation also require assessment. Establish the required admission time and signal stabilisation at the measurement point.
Medical, analytical, and life science applications
Mass spectrometry and calibrant introduction
Parker identifies calibrant gas control for mass spectrometers as a typical application. Series 9 can admit a reference gas or switch between connected routes. Qualify source conditions, connecting volume, leakage, and signal stabilisation within the instrument’s intended calibration and measurement sequence.
Gas chromatography and process analysis
Parker lists gas chromatography and process gas analysis among Series 9 applications. OEM designers can evaluate the valve for gas admission and route selection. Specify inlet and outlet conditions, then assess restrictions, purge behaviour, and switching effects on the complete analytical circuit.
Medical imaging and radiochemistry
Parker’s separate Series 9 liquid datasheet identifies radioactive liquids in medical imaging and illustrates a PET radiochemistry circuit. Assess the selected valve against the actual liquid chemistry, pressure, cleaning, and delivery conditions. Suitability must be established within the intended medical instrument.
Life science and OEM instrument assemblies
Designers can evaluate Series 9 for gas and compatible liquid switching within laboratory instrument modules. Match the shutoff or selection function to the operating sequence, then qualify wetted materials, leakage, flushing, and residual media under representative research or life science workflow conditions.
Why specify Parker Series 9 valves
Low leakage for analytical gas control
Parker’s gas performance table publishes a helium leak rate of 1 × 10⁻⁷ atm cc/sec. This gives designers a reference for defining leakage acceptance criteria. Assess the selected valve and assembled connections against the instrument’s vacuum conditions, background signal, and calibration requirements.
Fast switching with defined response times
Parker publishes response below 5 ms for the 0.030 and 0.060 inch orifices, and below 6 ms for the 0.116 inch version. These figures inform switching design and driver selection. Calibrant arrival and measurement stabilisation also depend on the source, restrictions, tubing, and connected volume.
Documented materials for chemical selection
Series 9 combines a 316 stainless steel body with configuration dependent FKM or FKM and Vespel sealing components. Other wetted materials include PTFE and stainless steel. Review the complete material arrangement against process media, cleaning agents, temperature, and exposure before selecting an OEM component.
Configurations for different circuit requirements
Two way normally closed and three way configurations provide shutoff and route selection functions. Orifice, pressure rating, coil voltage, and connections must be considered together. These documented choices help designers define the operating sequence and enable procurement teams to identify the required component accurately.
Series 9 specifications for OEM selection
Series 9 gives OEM teams documented operating limits and configuration choices for demanding analytical fluid control. These provide a practical basis for component selection, prototype qualification, and accurate production sourcing.
The following values summarise Parker’s published Series 9 gas datasheet. Confirm the current specification and drawing for the exact part, particularly where a custom configuration or liquid duty is required.
| Specification | Published family information |
|---|---|
| Valve type | Non-isolation solenoid valve |
| Functions | Two way normally closed or three way |
| Body material | 316 stainless steel |
| Sealing materials | FKM or FKM and Vespel, configuration dependent |
| Other wetted materials | PTFE, stainless steel, and FKM |
| Standard coil voltage | 12 or 24 VDC |
| Rated coil power | 12 W at either voltage in the referenced gas datasheet |
| Rated current | 1 A at 12 VDC; 0.5 A at 24 VDC |
| Ambient operating range | 4 to 105°C |
| Duty cycle | 100% continuous duty, within published operating conditions |
| Internal volume | 342.7 to 540.6 µL; confirm the selected configuration |
| Gas leakage specification | 1 × 10⁻⁷ atm cc/sec helium |
| Recommended filtration | 40 µm maximum |
| Porting | A-LOK compression fittings, 1/4-28, or 1/8 inch FNPT, configuration dependent |
Confirm the exact coil specification before designing the driver, as Parker’s product listing and linked datasheet show different power figures.
Internal valve volume is one part of the total connected volume. Tubing, fittings, restrictions, and chambers also affect flushing and the time required for a changed source to reach the measurement point.
Match orifice size to pressure and switching duty
Maximum operating pressure varies with both orifice size and valve function. A larger passage must be assessed against its own rating.
| Orifice diameter | Valve function | Maximum pressure | Response time |
|---|---|---|---|
| 0.030 inch / 0.76 mm | Two way NC | 1250 psig / 86.2 bar | Below 5 ms |
| 0.060 inch / 1.52 mm | Two way NC | 250 psig / 17.2 bar | Below 5 ms |
| 0.060 inch / 1.52 mm | Three way | 100 psig / 6.9 bar | Below 5 ms |
| 0.116 inch / 2.95 mm | Two way NC or three way | 100 psig / 6.9 bar | Below 6 ms |
Parker’s gas flow curves were measured with air at 24°C; the separate liquid curves use water at 24°C. Establish flow using the intended medium, pressure differential, and circuit arrangement. Confirm vacuum requirements against the selected part’s current specification.
The above component specifications come from Parker’s Series 9 Calibrant Valve datasheet.
Learn more about Parker Miniature Valves and Components
Integrating Series 9 into an OEM fluid circuit
Qualify the calibrant path and measurement response
Start with the source, required switching sequence, and pressure at each valve connection. Select the two way or three way function, then assess regulators, restrictions, tubing, fittings, and connected volume alongside the valve.
For gas calibration, measure arrival time, stabilisation, and residual signal at the analytical measurement point. For liquid duties, qualify flow and delivery with representative viscosity and temperature. Include leakage, filtration, and media compatibility throughout the assembled circuit when establishing the instrument’s operating and calibration sequence.
Match electrical drive to the operating sequence
For the datasheet’s 12 W coils, assess supply capacity and enclosure temperature against the number of valves energised together. Parker documents a 25 ms rated voltage hit followed by reduced holding voltage. Holding settings differ between two way and three way configurations.
Validate the driver, opening and release timing, and holding margin under representative pressure and temperature conditions. Check suppression components, electrical connections, and mounting space against the selected part’s current drawing before finalising the assembly and production specification.
Review every wetted material
Check the sealing components alongside the stainless steel body, PTFE coated armature, tubing, and fittings. Parker’s abbreviated liquid compatibility chart does not recommend FKM with methanol, acetonitrile, or tetrahydrofuran. A Vespel poppet does not remove FKM elsewhere in the wetted path.
Confirm suitability using the actual media, concentrations, temperatures, exposure periods, and cleaning sequence. For repeated source switching, assess residual media and the flushing time needed to meet the instrument’s acceptance criteria.
Series 9 ordering references
These examples come from Parker’s published gas ordering table. They are reference configurations; confirm the current part record, materials, pressure rating, and availability before ordering.
| Configuration | Voltage | Porting | Parker part number |
|---|---|---|---|
| 0.030 inch two way NC | 12 VDC | 1/16 inch A-LOK | 009-0100-900 |
| 0.030 inch two way NC | 24 VDC | 1/16 inch A-LOK | 009-0172-900 |
| 0.030 inch two way NC | 24 VDC | 1/4-28 | 009-0272-900 |
| 0.060 inch two way NC | 24 VDC | 1/8 inch A-LOK | 009-0270-900 |
| 0.060 inch two way NC | 24 VDC | 1/8 inch FNPT | 009-0631-900 |
| 0.060 inch three way | 24 VDC | 1/8 inch A-LOK | 009-0269-900 |
| 0.060 inch three way | 24 VDC | 1/8 inch FNPT | 009-0933-900 |
| 0.116 inch two way NC | 24 VDC | 1/8 inch FNPT | 009-0089-900 |
| 0.116 inch three way | 12 VDC | 1/8 inch FNPT | 009-0207-900 |
| 0.116 inch three way | 24 VDC | 1/8 inch FNPT | 009-0143-900 |
The 0.030 inch examples use Vespel and FKM sealing components; the listed larger orifices use FKM. Verify the complete wetted material arrangement for the selected part.
Specify with confidence from prototype to production
the right configuration for procurement
Specify the Parker part number alongside valve function, orifice, sealing materials, coil voltage, porting, and driver requirements. Include the actual gas or liquid duty, pressure conditions, leakage acceptance criteria, and mounting arrangement.
For development enquiries, provide sample quantities, calibrant or process media, flow and timing targets, and the qualification timetable. For production sourcing, include forecast volumes, delivery requirements, documentation, and agreed restrictions on substitutions.
Assess acquisition cost alongside driver hardware, assembly time, qualification, and maintenance. Confirm current lead times and availability for the selected configuration. Changes between Series 9, Series 99, or Pulse valves should undergo engineering review against the instrument’s requirements.
Discuss your OEM calibrant valve requirements
AEG’s OEM Fluid Technology Division combines practical experience in medical and analytical fluid control with knowledge of miniature valve specification, material compatibility, and system integration. Our team can guide component selection and assembly development, considering valve function, pressure differential, leakage, wetted materials, electrical drive, and calibrant delivery.
Share your fluidic schematic, calibrant or process media, inlet and outlet pressures, required flow, switching sequence, electrical supply, temperature conditions, and expected quantities so we can assess your requirements and recommend suitable options.
Browse the Parker Series 9 calibrant valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.
Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.
From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.
Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.