Parker Series 18 PTFE Manifold Valves for OEM Fluid Control

Labelled four station Parker Series 18 valve showing solenoid coils, the PTFE manifold body, threaded fluid ports, and mounting holes.

Compact multichannel liquid selection for medical, analytical, and life science instruments.

As a supplier of Parker precision fluid control products, Air Engineering Group also helps OEM teams specify valves and fluid control assemblies for medical, analytical, and life science applications.

Parker Series 18 valves are miniature liquid manifolds incorporating two, three, or four independently operated diaphragm isolation valve stations. Each station is a two port, normally closed valve, with all wetted parts made from PTFE.

The manifold selects liquids from multiple inlets into a common outlet, or routes one liquid supply to selected outlets. Its integrated passages therefore reduce the tubing and connections needed between separate valves. In addition, the published operating range is vacuum to 20 psig, approximately 1.4 bar.

How Parker Series 18 valves control liquids

Energising a station’s solenoid moves the internal PTFE poppet diaphragm to open its liquid path. Removing power allows the spring to return that station to its closed position. In addition, diaphragm isolation separates the liquid from the electromagnetic actuator.

The controller determines which channels open and for how long. Sequential operation can therefore select reagents or solvents; timed operation can contribute to mixing or gradient generation. Delivered proportions also depend on pressure, flow resistance, pump operation, and the surrounding fluidic circuit.

Electrical state of each stationLiquid path
Coil deenergisedNormally closed path closed.
Coil energisedSelected liquid path open.

Medical, analytical, and life science applications

OEM fluid control assemblies

Series 18 can be evaluated for OEM fluid control modules combining liquid selection with pumps, sensors, and instrument components. Integrating valve stations within one PTFE manifold also reduces interconnecting tubing. Therefore, designers should consider mounting space, port access, electrical drive, and flushing requirements when developing the assembly.

Medical and clinical diagnostic equipment

Potential OEM uses include selecting reagents, wash solutions, and cleaning liquids in diagnostic instruments. The PTFE wetted path is also relevant where chemical resistance and separation from the actuator are important. However, suitability must be established for the actual instrument and liquids.

Life science and laboratory automation

Designers can evaluate Series 18 for buffer selection and liquid routing between reservoirs and shared processing paths. These applications require attention to flushing, residual reagent, flow conditions, and the instrument’s qualification requirements.

Analytical instruments and chromatography

Parker identifies liquid chromatography gradient generation and liquid selection and mixing as typical applications. For example, Series 18 can select or proportion solvents on the low pressure side of a chromatography circuit, upstream of the high pressure pump.

Why specify Parker Series 18 valves?

An all PTFE wetted path

Parker specifies PTFE for both the body and seals. In addition, its abbreviated compatibility chart rates media including methanol, acetonitrile, tetrahydrofuran, toluene, and saline as excellent. However, check the full fluid chemistry, concentration, temperature, and cleaning regime when selecting the valve and connecting components.

Integrated, symmetrical flow paths

The shared manifold reduces interconnecting tubing and associated volume. Parker’s symmetrical design also gives each channel matched path length and timing, providing a useful basis for repeatable liquid selection and proportioning. However, the shared outlet passage still needs to be included in flushing and carryover assessments.

Documented switching and endurance

In addition, the published response time is below 12 ms, and the life rating is 10 million cycles, dependent on the application. Parker also reports testing beyond 10 million cycles with the valve subsequently meeting response, internal and external leakage, and repeatability specifications.

Evidence for OEM qualification

Parker’s published test results provide useful qualification evidence for OEM designers assessing switching response, internal and external leakage, and repeatability. However, actual service life remains dependent on the media and operating conditions within the instrument.

Published Series 18 specifications

Use these family specifications to shortlist a configuration, then confirm the selected part’s drawing and electrical requirements.

SpecificationPublished information
Valve functionDiaphragm isolation; two port, normally closed at each station.
Station optionsTwo, three, or four independently operated stations.
Media and wetted materialsLiquids; PTFE body and seals.
Operating pressureVacuum to 20 psig, approximately 1.4 bar.
Response timeBelow 12 ms.
LeakageBubble tight, as described by Parker.
Ambient temperature and duty4 to 66°C; 100% continuous duty rating.
Coil voltage and rated power9 or 12 VDC: 2.5 W; 24 VDC: 4.2 W per coil. Availability depends on configuration.
Fluid ports1/4-28 threaded, flat bottom ports.
Recommended filtration10 µm maximum.
Published life rating10 million cycles, application dependent.

The specification page lists a 0.060 inch / 1.52 mm orifice, while its ordering table lists 0.062 inch / 1.57 mm. Confirm the orifice against the exact part specification before finalising flow calculations.

The above component specifications come from Parker’s Series 18 Valve datasheet.

Learn more about Parker Miniature Valves and Components

Design the manifold into your fluidic circuit

Account for shared liquid volume

The valve’s electrical response and the liquid changeover measured at the instrument are different quantities. For example, tubing volume, viscosity, flow rate, and shared passages affect delivery delay and the flushing needed between liquids.

Parker lists the following internal volumes separately. The valve volume alone should therefore not be treated as the volume of the entire manifold.

StationsValve volumeInlet pathOutlet path
212.6 µL25.2 µL42.4 µL
312.6 µL25.2 µL42.4 µL
412.6 µL25.2 µL57.1 µL

Match electrical drive to the operating sequence

Assess power and enclosure temperature for the number of stations energised together. Parker documents a hit and hold method using 25 ms at rated voltage, followed by a reduced holding voltage: 5 V and 0.44 W for a 12 V coil, or 8 V and 0.46 W for a 24 V coil.

Validate the driver, opening and release timing, and holding margin under the intended pressure and temperature conditions. Define when channels may open together to manage mixing and unintended crossflow between connected supplies.

Qualify the wetted circuit for reliable operation

Check tubing, fittings, pumps, and cleaning agents for compatibility with the intended liquids and operating conditions. Test flow at representative pressure differentials, accounting for viscosity and connecting volume within the assembled circuit. Parker’s flow curve was measured with water at 24°C.

Include filtration, leakage, flushing, and switching performance in prototype qualification. Confirm that cleaning removes residual liquid from shared passages under representative operating conditions.

Specify the configuration with confidence

Achieving the right configuration for procurement

Provide the exact Parker part number, station count, coil voltage, port assignments, mounting arrangement, and driver requirements. Confirm the common port’s role and the selected orifice rather than ordering by the Series 18 family name alone.

For development enquiries, provide sample quantities, operating and cleaning liquids, inlet and outlet pressures, required flow, switching sequence, and qualification timetable. 

For production sourcing, include forecast volumes, delivery requirements, documentation, and agreed restrictions on substitutions.

Assess acquisition cost alongside fittings, driver hardware, assembly, qualification, and maintenance. Confirm current lead times and availability for the selected configuration.

Discuss your OEM gas pulse requirements

AEG’s OEM Fluid Technology Division combines practical experience in medical and analytical fluid control with detailed knowledge of miniature liquid valve specification, chemical compatibility, and manifold integration. Our team can guide component selection and assembly development, considering station count, shared fluid volume, operating pressures, electrical drive, and liquid switching performance.

Share your fluidic schematic, process and cleaning liquids, inlet and outlet pressures, required flow, switching sequence, electrical supply, flushing requirements, and expected quantities so we can assess your requirements and recommend suitable options.

Browse the Parker Series 18 valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.

Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.

From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.

Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.