Parker Series 11, 25, and 26 Miniature Solenoid Valves for OEMs

Parker miniature solenoid valves with labelled electrical leads, coil, valve body, male threaded connection, and female threaded port.

Pneumatic gas switching with manifold, barbed, and threaded connection options

Air Engineering Group supplies Parker Series 11, 25, and 26 miniature solenoid valves for OEM pneumatic assemblies. These direct acting poppet valves provide gas switching for medical, analytical, and general instrument circuits, with connection arrangements that suit different assembly layouts.

Select the series by its pneumatic interface, then choose the pressure model, valve function, material, voltage, and electrical termination. A larger orifice, lower power rating, or different connection changes the selection requirements. Assess inlet and outlet pressures, flow, leakage, and operating duty together.

The range provides options for connecting tubing, mounting onto a manifold, or using threaded body ports. Include fittings, seals, wiring, and service clearance in the device layout, then qualify the assembled circuit for pressure response, switching, leakage, and temperature.

How these miniature solenoid valves operate

Electrical energisation moves a poppet to open, close, or redirect the gas path. Two port versions control a single path, while three port versions connect a common port to alternative paths. Choose the function around the required energised and de energised circuit states.

Parker provides pressure and vacuum porting diagrams. Follow the drawing for the selected function when assigning supply, load, exhaust, and vacuum connections. Verify the operating sequence during power interruption, start up, and transitions between states.

Choosing between Series 11, 25, and 26

The body and stem connections are separate ordering choices. Confirm both against the assembly drawing, including tubing size and any restrictions associated with the orifice or function. The published charts restrict the manifold stem option to specified normally closed configurations.

Use the required connection to guide mechanical integration. Assess sealing, fastening, fitting access, and tube routing before finalising the series. Match the mounting arrangement to the permitted loads and replacement method within the device.

SeriesBody connectionBody material options
Series 11Manifold mount or tubing barbsBrass or 303 stainless steel
Series 2510-32 female threaded body portsNickel plated brass
Series 2610-32 male threaded body connectionNickel plated brass

Parker Series 11, 25, and 26 Solenoid valve OEM applications

Oxygen concentrators and conservers

Define gas composition, oxygen concentration, pressure, flow, and switching duty. Confirm the required cleaning specification and compatible materials for the selected assembly. For bed switching or delivery circuits, qualify pressure recovery, leakage, and power loss behaviour with the intended filters and connected volumes.

Compression therapy equipment

Assess inflation, isolation, and venting against garment volume, tubing resistance, and the required pressure profile. Choose the function for each circuit state, then qualify leakage and pressure response. Include simultaneous channel operation, repeated switching, power interruption, and enclosure temperature.

Gas chromatography and analytical instruments

Assess materials, cleanliness, leakage, and residual gas against the measurement method. Define sample arrival, purge, and settling targets at the detector. Test the complete gas path with the intended gas, manifold, tubing, and operating temperatures rather than relying on the valve switching reference alone.

Insufflation and medical gas equipment

Evaluate the valve within the device’s gas supply, isolation, or venting circuit. Define pressure limits, flow demand, contamination protection, and behaviour during loss of power. Qualify the assembled instrument with its regulation, monitoring, filters, and intended operating sequence under representative conditions.

Parker Series 11, 25, and 26 Valve Selection Benefits

Connection choices for assembly integration

The three body interfaces allow selection around the tubing or manifold arrangement. Compare the installation envelope, sealing method, fitting access, and service requirements. Include volume and restrictions introduced by external connections when assessing circuit response.

Material selection for the gas path

Series 11 offers brass or stainless steel body options, while Series 25 and 26 use nickel plated brass bodies. Assess the selected wetted construction and FKM seals against gas composition, temperature, cleanliness, and the measurement requirements of the device.

Electrical termination options

Wire leads, PCB pins, solder tabs, and quick connect spades provide alternative electrical interfaces. Select the termination to match the PCB or harness, assembly process, strain relief, and replacement access. Confirm the coil voltage and power alongside the connection choice.

Model specific pressure and power selection

The model table links orifice size, nominal Cv, maximum supply pressure, power, and life requirements. Compare the complete row when selecting a valve. Validate timing and service life using the intended gas, electrical drive, connected circuit, and operating duty.

Valve Technical specifications

Confirm function availability in the selected series’ ordering chart. The Series 26 chart lists two port normally closed, three port normally closed or distributor, and three port normally open configurations. Do not assume every function is available with every connection.

ParameterPublished specification
Valve designDirect acting poppet
FunctionsTwo or three port normally closed, normally open, or distributor, according to ordering configuration
MediaNon reactive gases; confirm compatibility and service requirements
Voltage options5, 12, or 24 VDC
Power options0.5, 1.0, or 2.0 W, according to model
Standard nominal orifices0.76 or 1.27 mm
Additional Series 11 orifice2.59 mm, Model 20 only
Maximum range pressureUp to 100 psig; individual model limits apply
Published vacuum rangeUp to 27 inHg, 686 mmHg vacuum
Cycling responseLess than 30 ms
Published air leakageLess than 0.016 sccm; Series 11 Model 20 reference is 1.0 sccm
Operating environment0°C to 70°C
Storage temperatureMinus 40°C to 70°C
Recommended filtration40 µm
Published weight60 g
Published internal volume0.426 cm³

Standard model performance parameters

These seven model references appear across all three series. Bar equivalents are rounded to two decimal places. Confirm the permitted inlet, outlet, and differential pressures for the selected function. The vacuum figure is a vacuum reading rather than an absolute pressure value.

Parker’s life requirements vary by model and depend on elastomer, duty cycle, and pressure. The 260 million cycle reference applies to Models 16 and 18. Use the selected model’s figures as qualification inputs, then validate timing and service life within the instrument.

ModelNominal orificeNominal CvMaximum supply pressurePowerLife requirement in million cycles
100.76 mm0.017100 psi / 6.89 bar2.0 W100
121.27 mm0.03150 psi / 3.45 bar2.0 W100
130.76 mm0.01750 psi / 3.45 bar1.0 W200
151.27 mm0.02525 psi / 1.72 bar1.0 W200
160.76 mm0.01725 psi / 1.72 bar0.5 W260
181.27 mm0.02110 psi / 0.69 bar0.5 W260
191.27 mm0.02570 psi / 4.83 bar2.0 W20

The above component specifications come from Parker’s Series 11/25/26 Miniature Solenoid Valve datasheet.

Learn more about Parker Miniature Valves and Components

Series 11 Model 20

Model 20 has a 2.59 mm nominal orifice, nominal Cv of 0.069, 25 psi maximum supply pressure, 1.0 W power rating, and a 25 million cycle life requirement. Parker lists an air leakage reference of 1.0 sccm for this model.

The ordering chart restricts Model 20 to the two port normally closed universal function, brass with FKM, and a manifold mounted body. Confirm the porting and current drawing before using this configuration in a circuit.

Flow and connected volume

Parker’s typical flow curves use air at 24°C. Use the curve for the selected model and state the gas and reference conditions when specifying flow. Evaluate inlet and outlet pressures together, especially where the instrument operates against variable back pressure or vacuum.

Measure circuit performance with the intended filters, tubing, fittings, and manifold. Valve flow and switching references do not determine sample arrival, inflation, venting, or evacuation time for the connected device volume.

Wetted materials and gas compatibility

Use the current drawing for manifold port positions, sealing features, fasteners, and the selected threaded or barbed interface. Check fitting loads, tube retention, seal compression, and access for assembly and replacement. Verify port isolation and leakage after mounting.

Allow space for the complete installed assembly. The published dimensions depend on the series and connection arrangement, and the outline drawings include additional interface details. Confirm the installation envelope before releasing manifold or enclosure drawings.

Parker Series 11/25/26 Manifold and electrical integration

Mounting and sealing

Use the current drawing for manifold port positions, sealing features, fasteners, and the selected threaded or barbed interface. Check fitting loads, tube retention, seal compression, and access for assembly and replacement. Verify port isolation and leakage after mounting.

Allow space for the complete installed assembly, including service access. The published dimensions depend on the series and connection arrangement, and the outline drawings include additional interface details. Confirm the installation envelope before releasing manifold or enclosure drawings.

Coil power and reduced power drive

Assess supply capacity and thermal loading using the selected coil rating, simultaneous valve operation, and duty. Include wiring losses and voltage at the coil, then measure enclosure temperature during prototype qualification.

Parker identifies spike and hold or PWM as possible methods of reducing power. Obtain drive guidance for the selected configuration and validate opening, holding margin, release timing, suppression, and temperature with the actual driver. Confirm numerical settings rather than transferring them from another valve family.

Qualification and production planning

the right configuration for procurement

Qualify flow, leakage, switching, pressure response, and temperature with the intended circuit. Test simultaneous operation and power interruption. Include application targets for sample transit, pressure recovery, venting, or vacuum performance.

Specify the series, pressure model, function, material, voltage, electrical termination, and both pneumatic connection codes. Provide gas composition, inlet and outlet pressures, flow, temperatures, filtration, connected volume, and operating duty.

Confirm the drawing, porting, seals, and accessories. Provide development quantities and qualification timescales, plus production forecasts, delivery requirements, and documentation. Assess assembly and lifecycle costs, and review substitutions or configuration changes with engineering.

Discuss your OEM gas control application

AEG’s OEM Fluid Technology Division can guide component selection and assembly development, considering gas compatibility, pressure, vacuum, flow, valve function, electrical drive, manifold sealing, filtration, and maintenance requirements.

Share your schematic, gas composition, inlet and outlet pressures, required flow, switching sequence, electrical supply, temperatures, connected volume, and quantities. Include sample timing, venting, or vacuum targets so we can assess suitable valve configurations and integration options.

Browse the Parker Series 11, 25, and 26 valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.

Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.

From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.

Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.