Parker Series 1 and 2 PTFE Isolation Valves for OEMs

Labelled Parker Series 1 PTFE isolation valve showing electrical leads, solenoid actuator, PTFE valve body, and threaded fluid ports.

Miniature diaphragm isolation valves for reagent control, solvent management, and OEM liquid circuits

As a supplier of Parker precision fluid control products, Air Engineering Group brings practical experience in specifying miniature valves for medical and analytical OEM fluid control assemblies.

Parker Series 1 and 2 are miniature solenoid operated diaphragm isolation valves for liquid control. The diaphragm separates the wetted flow path from the electromagnetic actuator, allowing liquids to pass through chemically resistant materials while the solenoid operates the switching mechanism.

Series 1 provides three way liquid selection or diversion, while Series 2 provides two way normally closed shutoff. Both use PTFE wetted bodies and diaphragms, with borosilicate glass also present in the Series 1 wetted path. In addition, their published pressure range extends from vacuum to 20 psig, approximately 1.4 bar.

How Parker Series 1 and 2 isolation valves work

Energising the coil moves the internal mechanism and changes the diaphragm’s position at the valve seat. Series 2 opens its normally closed flow path. In contrast, Series 1 switches its common port between the normally open and normally closed connections, allowing a circuit to select between two routes.

Both valves are direct acting and can actuate without applied pressure or vacuum. However, liquid flow still requires a suitable driving pressure or vacuum. Delivered volume depends on pressure differential, viscosity, valve timing, and the connected circuit, so timed switching should be qualified within the assembled instrument.

Electrical stateSeries 1: three waySeries 2: two way normally closed
Coil deenergisedCommon port 2 connects to normally open port 3; port 1 is closed.Flow path closed.
Coil energisedCommon port 2 connects to normally closed port 1; port 3 is closed.Flow path open.

Use Parker’s port diagram and the selected part drawing when assigning connections. Define the required fluid path during power loss as well as during normal operation.

Medical, analytical, and life science applications

Medical and clinical diagnostic equipment

OEM designers can evaluate these valves for admitting reagents, selecting wash liquids, and isolating liquid lines in diagnostic instruments. In addition, diaphragm isolation separates the liquid path from the actuator. However, qualify the selected configuration against reagent chemistry, cleaning procedures, and the instrument’s performance requirements.

Analytical instruments and solvent handling

Parker identifies reagent control, solvent management, and aggressive liquid control as typical applications. Series 1 can select between liquid routes, while Series 2 can admit or isolate a supply. Therefore, size the circuit using the intended liquid and pressure differential, within the published operating range.

Life science and laboratory automation

Potential uses include buffer selection, wash liquid routing, and reagent switching within automated laboratory equipment. The defined internal volume also helps designers plan flushing and residual liquid control. Then evaluate connecting tubing, switching sequences, and cleaning effectiveness using the actual liquids and representative instrument conditions.

OEM liquid control assemblies

Individual Series 1 and 2 valves can be incorporated into assemblies combining pumps, reservoirs, sensors, and fluid connections. Their separate switching points also allow designers to distribute liquid control around an instrument. Therefore, consider mounting access, fitting space, electrical connections, and service requirements when developing the assembly.

Why specify Parker Series 1 and 2 isolation valves?

Isolation for demanding liquid chemistry

PTFE bodies and diaphragms provide a chemically resistant wetted path separated from the solenoid actuator. Series 1 also includes borosilicate glass, which must be considered during compatibility assessment. This documented material arrangement helps designers evaluate both process liquids and cleaning agents before committing to a configuration.

Two configurations for different fluid paths

Series 1 three way selection and Series 2 normally closed shutoff address different circuit requirements. Designers can specify the route connected or isolated in each electrical state. Choosing the correct function early helps establish the operating sequence, port assignment, and behaviour when electrical power is removed.

Defined volume and switching performance

Parker publishes internal volumes of 96 µL for Series 1 and 49 µL for Series 2, with cycling responses below 12 ms and 20 ms respectively. These values inform fluid circuit planning and prototype assessment. Include fittings and tubing when evaluating total retained liquid and switching performance.

Evidence for OEM qualification

The published life rating is 10 million cycles and is application dependent. Parker also specifies continuous duty operation and bubble tight leakage performance. These ratings provide useful starting points for qualification, while actual service life and instrument performance depend on media, contamination, electrical drive, and operating conditions.

Series 1 and 2 isolation valve specifications

The family offers documented materials, switching functions, and operating specifications for OEM liquid control. These give designers a basis for component selection and prototype testing, while helping procurement teams identify the required configuration accurately.

The following values are published in Parker’s Series 1 and 2 datasheet. Confirm the selected part’s current specification and drawing before finalising an OEM design.

SpecificationSeries 1Series 2
Valve configurationThree way selectorTwo way, normally closed
Wetted body and diaphragmPTFEPTFE
Additional wetted materialBorosilicate glassNone listed
Operating pressureVacuum to 20 psig, approximately 1.4 barVacuum to 20 psig, approximately 1.4 bar
Orifice diameter0.060 inch / 1.52 mm0.060 inch / 1.52 mm
Internal volume96 µL49 µL
Published cycling responseLess than 12 msLess than 20 ms
Standard coil voltages12 or 24 VDC12 or 24 VDC
Published rated coil power2.5 W at 12 VDC; 4.2 W at 24 VDC2.5 W at 12 VDC; 4.2 W at 24 VDC
Ambient operating temperature4 to 66°C4 to 66°C
Duty rating100% continuous duty100% continuous duty
Fluid connections1/4-28 threaded ports1/4-28 threaded ports
Published leakage performanceBubble tightBubble tight
Published life rating10 million cycles, application dependent10 million cycles, application dependent

Parker’s typical flow curve was measured with water at 24°C. Use it as a reference for sizing, then validate flow with the intended liquids, temperatures, and pressure differentials. The datasheet gives 10 µm filtration guidance; confirm the appropriate filter specification for the application.

The above component specifications come from Parker’s Series 1 and 2 Isolation Valve datasheet.

Learn more about Parker Miniature Valves and Components

Integrate the valve into your OEM instrument

Specify compatibility across the wetted circuit

Select the valve against every process liquid, reagent, solvent, and cleaning agent. Series 2’s listed wetted materials are PTFE, while Series 1 also includes borosilicate glass. Parker’s abbreviated chart does not recommend borosilicate glass with toluene or 20% sodium hydroxide, despite rating PTFE as excellent with both.

Review tubing, fittings, pumps, reservoirs, and seals alongside the valve. Confirm suitability at the intended concentrations, temperatures, and exposure times. Include flushing, leakage, residual liquid, and switching performance in prototype qualification, accounting for viscosity and the volume between the valve and the measurement or delivery point.

Define the fluid path and mechanical layout

Start with the fluidic schematic and required switching sequence. For Series 1, identify the common, normally open, and normally closed connections, including the route maintained during power loss. For Series 2, confirm where normally closed isolation is required.

Check mounting, fitting geometry, tubing access, and electrical clearance against the selected part drawing. Confirm supplied lead length before finalising the harness. Keep connecting passages short where practical, then validate flow, flushing, and residual liquid in the assembled circuit.

Match electrical drive to the operating sequence

Assess supply capacity and enclosure temperature for the number of valves energised together. Parker documents a hit and hold method using 25 ms at rated voltage, followed by the reduced holding settings below. Series 1 and Series 2 require different holding voltages.

Validate the driver, opening and release timing, and holding margin under representative pressure and temperature conditions. Coil suppression can affect release timing, so check the resulting liquid switching at the instrument’s measurement point during prototype qualification.

Valve and rated voltageHold voltagePublished hold power
Series 1, 12 VDC6 V0.63 W
Series 1, 24 VDC12 V1.04 W
Series 2, 12 VDC5 V0.44 W
Series 2, 24 VDC8 V0.46 W

These are Parker’s published holding settings for the respective configurations.

Parker's Published standard part numbers

Parker lists the following standard configurations. Confirm current availability and any project specific requirements when ordering.

SeriesFunctionCoil voltageParker part number
Series 1Three way12 VDC001-0017-900
Series 1Three way24 VDC001-0028-900
Series 2Two way, normally closed12 VDC002-0017-900
Series 2Two way, normally closed24 VDC

002-0010-900

The part numbers identify the published configurations; they do not indicate current stock availability.

Specify with confidence from prototype to production

the right configuration for procurement

Specify the exact Parker part number alongside series, switching function, coil voltage, porting, and mounting requirements. Confirm the supplied lead length and selected part drawing. A three way Series 1 valve and a two way Series 2 valve perform different functions and require engineering review before substitution.

For development enquiries, provide sample quantities, process and cleaning liquids, inlet and outlet pressures, required flow, switching sequence, and the qualification timetable. For production sourcing, include forecast volumes, delivery requirements, documentation, and any agreed restrictions on substitutions.

Assess acquisition cost alongside driver hardware, fittings, assembly time, qualification, and maintenance. Confirm lead times and current availability for the selected part, as these can vary between configurations.

Discuss your OEM isolation valve requirements

AEG’s OEM Fluid Technology Division combines practical experience in medical and analytical fluid control with detailed knowledge of miniature isolation valve specification, chemical compatibility, and system integration. Our team can guide component selection and assembly development, considering switching function, wetted materials, pressure conditions, connecting volume, and electrical drive.

Share your fluidic schematic, process and cleaning liquids, inlet and outlet pressures, required flow, switching sequence, electrical supply, flushing requirements, and expected quantities so we can assess your requirements and recommend suitable options.

Browse the Parker Series 1 and 2 valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.

Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.

From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.

Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.