Parker PND Series Miniature Pneumatic Solenoid Valve Applications

Labelled Parker PND pneumatic solenoid valve showing the coil, electrical connection, tubing barb, and metal frame.

Normally open exhaust control for OEM medical and pneumatic equipment

The Parker PND Series is a miniature, two port, normally open pneumatic solenoid valve with a direct acting poppet mechanism for releasing pressure from low pressure gas circuits. With no electrical power, the exhaust path is open to atmosphere. Energising the coil closes that path, allowing the connected circuit to retain pressure.

Parker PND valve applications include non invasive blood pressure devices and other medical pneumatic systems. The series gives OEM designers a compact exhaust component with defined pressure, power, leakage, and operating specifications. Its published operating range is 0 to 6 psig, approximately 0 to 0.4 bar. Valve selection should account for orifice size, tubing restrictions, and the required pressure release time.

As a supplier of Parker precision fluid control products, Air Engineering Group brings practical experience in specifying and integrating their valves into medical OEM solutions.

How does a Parker PND valve work?

The PND uses a direct acting poppet mechanism. Switching the coil power changes the state of the exhaust connection. In a blood pressure monitor, this can provide a vent path after measurement or when power is removed.

Electrical stateValve function
Coil energisedThe exhaust path closes, allowing the circuit to retain pressure.
Coil deenergised

The exhaust path opens, allowing the connected circuit to vent to atmosphere.

Actual pressure release depends on the starting pressure, cuff or chamber volume, tubing, and other restrictions. Designers should verify the response of the assembled circuit, including its behaviour when electrical power is lost.

Parker PND valve applications in medical equipment

Non invasive blood pressure monitoring is the clearest documented PND application. Parker also identifies patient monitoring, patient therapy, respiratory, and anaesthesia equipment as markets for the series.

Medical applicationRole to evaluate
Blood pressure monitorsCuff exhaust after measurement and an open vent path when power is removed.
Patient monitoring and therapy equipmentPressure release from suitable low pressure pneumatic circuits.
Respiratory and anaesthesia equipmentExhaust control in compatible low pressure pneumatic subsystems, subject to the device’s gas and performance requirements.

In analytical and life science equipment, designers can also evaluate PND valves for venting small gas chambers or auxiliary pneumatic circuits. Selection should follow the actual gas, pressure, and exhaust requirements; the PND specification covers gases rather than liquid reagents or solvents.

Why specify a Parker PND valve?

Low internal volume for faster fluid changes

Parker publishes model specific pneumatic, electrical, and mechanical data, helping OEM teams compare variants before prototyping. The normally open exhaust function, miniature format, and low published power consumption are useful starting points for compact medical equipment.

The PND-05A combines a larger orifice and higher nominal Cv with lower published power consumption. The PND-05D has a lower typical weight and smaller internal volume. Evaluate these differences against the required exhaust performance and available installation space.
Parker lists valve dimensions of 25.7 × 10 × 12 mm for the PND-05D. Use the drawing for the selected model when checking clearances, tubing routes, and electrical connections.

Choose between two defined model configurations:

SpecificationPND-05APND-05D
Orifice diameter1.27 mm / 0.050 inch0.76 mm / 0.030 inch
Nominal Cv0.0350.017
Published power0.36 W0.50 W
Typical weight17.0 g11.4 g
Internal volume0.056 cm³0.041 cm³

Manage power during the operating sequence

The coil requires power to keep the exhaust closed. Parker notes that pulse width modulation can reduce power consumption, which may help battery powered equipment. Verify actuation, holding, and release with the intended driver, supply voltage, duty cycle, and operating environment.

Use the published life rating in qualification

The published life rating is 250,000 cycles, with Parker reporting no performance degradation at that test point. This provides a useful reference for repeated exhaust operation. Expected service life still depends on duty cycle, gas quality, contamination, operating conditions, and maintenance.

Select the right PND valve for your circuit

Start with the initial pressure, gas volume, operating sequence, and required pressure release time. Then select the model and electrical configuration, accounting for tubing restrictions and the conditions inside the finished instrument.

For medical devices, qualification should include the intended use, pressure release behaviour, leakage, electrical control, and relevant failure conditions. Component specifications provide inputs to that work; suitability is established through the finished device’s design and validation.

ParameterPublished specificationDesign consideration
Pressure0 to 6 psig, approximately 0 to 0.4 barCheck the full operating range and possible pressure excursions.
MediaAir, nitrogen, argon, carbon dioxide, and other non-reacting gasesConfirm compatibility with the actual gas and wetted materials.
Coil voltage3, 6, or 12 VDCMatch the part number to the supply and driver.
Valve cycling responseLess than 100 msMeasure the required pressure release time in the assembled circuit.
Air leakageLess than 0.016 sccmAccount for closed valve leakage when assessing pressure retention.
Tubing connectionSingle barb for 2.0 mm ID tubingCheck tubing fit, routing, and pressure losses.
Operating environment0 to 55 °CCheck the temperature inside the equipment enclosure.
Filtration40 µm recommendedDefine contamination control for the pneumatic circuit.

The above component specifications come from Parker’s PND Valve datasheet.

Learn more about Parker Miniature Valves and Components

Specify with confidence from prototype to production

What procurement teams should confirm

Specify the exact Parker part number, including the PND-05A or PND-05D model and voltage suffix. For example, PND-05A-06 identifies the 6 VDC PND-05A configuration. Confirm the mechanical drawing and connection requirements before ordering.

For development enquiries, provide sample quantities, gas, pressure conditions, exhaust timing targets, and the qualification timetable. For production sourcing, include forecast volumes, delivery requirements, documentation, and any agreed restrictions on component substitutions.

Assess acquisition cost alongside driver requirements, assembly time, device qualification, and expected maintenance.

Confirm lead times and current availability for the selected part, as these can vary between configurations.

Discuss your OEM PND valve requirements

AEG’s OEM Fluid Technology Division combines practical experience in medical and analytical fluid control with detailed knowledge of miniature valve specification, pneumatic circuit design, and system integration. Our team can guide component selection and assembly development, considering operating pressure, orifice size, coil voltage, gas compatibility, and exhaust performance.

Share your pneumatic schematic, gas, operating conditions, cuff or chamber volume, electrical supply, required pressure release time, and expected quantities so we can assess your requirements and recommend suitable options.

Browse the Parker PND Series valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.

Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.

From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.

Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.