Parker C7 Cartridge Solenoid Valves for OEM Gas Control

Parker C7 pneumatic cartridge valve with labelled electrical leads, cartridge body, external O-ring seals, and gas openings

Compact pneumatic switching for portable instruments and medical devices

Air Engineering Group supplies Parker C7 miniature cartridge solenoid valves for OEM gas circuits where space, weight, and electrical power need careful management. The range offers two way normally closed and three way normally closed configurations for gas isolation or switching between circuit paths within a manifold.

Select the valve function, orifice, seal material, and coil together. Pressure capability and power consumption vary across the range, so assess the full part number against the required gas flow, pressure conditions, switching sequence, and available supply.

This guide covers pneumatic selection and integration. For liquid circuits, refer to the dedicated Parker C7 liquid valve information and confirm the required media compatibility and configuration.

What cartridge mounting provides

The C7 fits into a prepared manifold cavity, with external seals separating the gas passages. Positioning the valve close to the controlled volume can shorten tubing runs and reduce the volume between switching and measurement points. Include the manifold passages when assessing the resulting circuit response.

Plan the cavity, retaining hardware, wiring, and service access as one assembly. Control machining, deburring, and cleaning before insertion, then verify seal integrity and leakage. A compact valve requires sufficient access for assembly and replacement within the finished device.

How the C7 valve functions

The two way normally closed valve blocks its controlled passage when de-energised and opens when energised. Three way versions switch between the connections shown in Parker’s porting diagram. Specify the role of each port and its connection in both electrical states.

Parker provides separate arrangements for pressure and vacuum operation. Use the drawing for the selected configuration to define supply and controlled circuit connections. Verify behaviour during power loss, start-up, and switching transitions under representative operating conditions.

Parker C7 Cartridge Solenoid valve OEM applications

Portable respiratory equipment

Assess the gas path against pressure requirements, connected volume, and battery capacity. Qualify switching with the intended driver, including supply variation and repeated operation. Measure pressure response at the device’s controlled point and include simultaneous valve operation in thermal testing.

Air over liquid dispensing

The valve can switch the gas circuit used to pressurise or vent a dispense reservoir. Establish dosing performance with the actual liquid, reservoir headspace, pressure regulation, and outlet restrictions. Test changes in reservoir fill level and define protection against liquid migration into the pneumatic path.

Negative pressure wound therapy

Select the valve against the required vacuum level, leakage limit, venting sequence, and recovery time. Protect the pneumatic assembly from collected liquids and contamination. Validate performance with the intended filters, tubing, and connected volume under representative operating conditions.

Gas sampling and environmental monitoring

For sidestream CO₂ or environmental sampling assemblies, assess leakage, purge volume, moisture exposure, and sample transit time. Position switching components with the complete sample path in mind. Validate the arrival of the selected gas and the clearance of residual gas at the measurement point.

Parker C7 Cartridge Solenoid Valve Advantages

Integration within limited space

The miniature cartridge provides an option for closely grouped pneumatic channels. Evaluate the manifold footprint alongside retention, lead routing, filter access, and service clearance to determine the finished assembly size.

Configuration choice

Choose an orifice using the required operating point and the pressure limit for its valve function. Check that filters, fittings, and manifold passages allow the intended circuit performance, including venting or vacuum recovery where applicable.

Electrical power planning

The range includes lower power two way configurations. Use the selected coil’s published rating to estimate electrical demand and enclosure heat, then verify both under the intended switching duty and ambient conditions.

Performance qualification

Use Parker’s response and reliability references to set development targets. Validate device timing and service life with the actual gas, driver, pressures, temperatures, and operating sequence, including foreseeable supply variation.

Parker C7 Technical specifications

Published specifications from Parker’s C7 technical literature.

ParameterPublished specification
Valve functionsTwo way NC and three way NC
Cartridge dimensions7 mm diameter, 20 mm length
Weight3.1 g
Nominal orifices0.3, 0.5, 0.8, and 1.0 mm
Standard supply12 or 24 VDC, ±5%
Electrical connection80 mm flying leads, 28 AWG
Wetted body materialStainless steel Series 300 and 400
Internal and external sealsFKM or EPDM
Operating environment0°C to 50°C
Storage environment−40°C to 70°C
Maximum cycling response10 ms
Reliability referenceTwo way: 130 million cycles; three way: 55 million cycles
Reliability conditions0.90 reliability factor at 95% confidence
Internal valve volumeTwo way: 81 µL; three way: 90 µL

Pressure limits and filtration

Parker specifies these limits and filtration recommendations by configuration.

Nominal orificeTwo way maximum pressureThree way maximum pressureRecommended filtration
0.3 mm10 bar / 145 psi10 bar / 145 psi5 µm
0.5 mm8 bar / 116 psi6 bar / 87 psi10 µm
0.8 mm5 bar / 73 psi2.5 bar / 36.3 psi10 µm
1.0 mm3 bar / 43.5 psi1.5 bar / 21.8 psi10 µm

Use the limit for the selected function and orifice. For vacuum service, confirm absolute pressures, pressure relationships, and port assignments. The published pressure table does not establish a separate numerical ultimate vacuum rating.

Include filter pressure loss and contamination loading in circuit testing. Specify the maintenance interval and access needed to preserve the required flow as the filter accumulates particles.

Electrical requirements

Assess supply capacity and thermal loading against coil rating, simultaneous operation, and duty. Measure enclosure temperature during prototype testing.

Configuration12 VDC power24 VDC powerResistance at 20°C, 12 / 24 VDC
0.3 mm two way0.5 W0.6 W288 / 995 Ω
0.5 mm two way1.0 W0.85 W140 / 700 Ω
0.8 and 1.0 mm two way, plus all three way versions1.0 W1.2 W140 / 495 Ω

Published resistance tolerance is ±5%.

Materials and gas compatibility

Review the stainless steel body, internal seals, external seals, manifold, and any assembly lubricant against the gas and anticipated contamination. Choose FKM or EPDM using compatibility information for the specific exposure, temperature, and service conditions.

Include humidity, condensate, cleaning residues, and storage in the assessment. For oxygen containing circuits, confirm materials and cleanliness for the actual concentration and pressure. Verify leakage after cleaning and conditioning the assembly as intended for production.

The above component specifications come from Parker’s C7 Miniature Cartridge Solenoid Valve datasheet.

Learn more about Parker Miniature Valves and Components

Parker C7 Manifold and electrical integration

C7 Cartridge cavity and retention

Follow the current Parker cavity and mounting drawings. The published maximum insertion force is 30 N, with retention at the body indent. Design the retaining hardware to react operating loads through this feature and provide access for removal.

Control cavity finish, entry geometry, deburring, and cleanliness to protect the external seals. Confirm any insertion lubricant’s compatibility. Include retention inspection and circuit leak testing in the assembly process, with checks for the pressure and vacuum conditions the device will encounter.

C7 Hit and hold control

Parker’s optional guidance specifies at least 50 ms at rated voltage before reduced DC holding at 50% rated voltage. For PWM holding, it gives a minimum frequency of 1 kHz and a nominal 50% duty cycle.

Reduced DC voltage and PWM duty cycle describe different drive methods. Validate the selected driver for opening, holding margin, and release timing across the required pressures and temperatures. Include suppression components, supply variation, and simultaneous coil operation in the assessment.

Prototype qualification and production selection

the right configuration for procurement

Qualify flow, leakage, switching, pressure, and temperature with the intended circuit. Test simultaneous operation and power loss under representative conditions.

Specify the valve configuration, gas composition, pressures, flow, temperatures, filtration, manifold, and duty.

Parker lists C07-312EP03FF-000 as a 12 VDC gas valve with a 0.3 mm nominal orifice. Confirm the code, drawing, and accessories. The separate C07-MCS manifold includes a clip and screw.

Provide quantities, qualification timescales, forecasts, delivery needs, and documentation. Assess lifecycle costs and review changes with engineering.

Discuss your OEM gas control application

AEG’s OEM Fluid Technology Division can guide component selection and assembly development, considering gas compatibility, pressure differential, flow, electrical drive, manifold integration, filtration, and maintenance requirements.

Share your schematic, gas composition, inlet and outlet pressures, required flow, connected volume, switching sequence, electrical supply, temperatures, and quantities. Include timing, venting, or vacuum targets so we can assess suitable valve configurations and integration options.

Browse the Parker C7 pneumatic valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.

Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.