Parker C15 liquid valves are 15 mm diameter, manifold mounted solenoid cartridges for switching liquid flow. Parker’s liquid datasheet describes a two port, normally closed configuration with stainless steel wetted components and FKM or EPDM seals.
The cartridge fits into a compatible manifold cavity, where external O-rings seal the liquid connections. In addition, orifice sizes from 0.5 to 2.0 mm provide different flow and pressure capabilities, with the maximum operating pressure dependent on the selected orifice.
How Parker C15 liquid valves work
The electromagnetic coil actuates the internal plunger to open the normally closed liquid path. When electrical power is removed, the valve returns to its closed state. In addition, the surrounding manifold provides the connecting passages to the rest of the instrument.
Timed switching controls when liquid can flow. Delivered volume also depends on pressure differential, viscosity, orifice size, and circuit resistance. The published cycling response should therefore be assessed alongside the liquid delivery measured at the intended point of use.
| Electrical state | Valve function |
|---|---|
| Coil deenergised | Liquid path closed. |
| Coil energised | Liquid path open. |
Medical, analytical, and life science applications
Medical and clinical diagnostic equipment
Parker identifies clinical diagnostics as a market and reagent addition, wash, and waste as typical applications. OEM designers can therefore evaluate the C15 for these liquid switching functions, selecting the seals, orifice, and operating pressure to suit the instrument’s process and cleaning requirements.
Analytical instruments and environmental monitoring
The C15 is relevant to analytical liquid circuits that admit reagents, control washing, or route waste. Parker also identifies environmental monitoring as a market. Therefore, specify required flow at the intended pressure differential and verify material compatibility with the samples, reagents, and cleaning liquids used.
Life science and laboratory automation
Designers can evaluate C15 cartridges for liquid handling modules within automated laboratory and life science instruments. For example, potential uses include reagent admission and wash control. However, qualification should address switching consistency, fluid compatibility, residual liquid, and the maintenance requirements of the assembled circuit.
OEM fluid control assemblies
The cartridge format allows valves to be integrated alongside pumps, sensors, and other components within a custom manifold. This can therefore simplify connecting passages and assembly. However, bore geometry, seal access, retaining clip position, and electrical routing should be considered together when developing the module.
Why specify Parker C15 cartridge liquid valves?
Compact manifold integration
A published diameter of 15 mm, length of 29 mm, and weight of 22 g give OEM designers a defined component envelope. In addition, the cartridge and mechanical securing system allow the manifold to form the liquid connections, reducing the space needed for separately connected valves.
Electrical options for instrument integration
Standard 12 and 24 VDC variants provide options for different instrument supplies. In addition, typical rated power ranges from 1.1 to 1.7 W, depending on orifice and voltage. Parker also documents an optional hit and hold method for reducing holding power and heat generation.
Evidence for OEM qualification
Parker publishes a 500 million cycle rating for the two way valve, with a 0.90 reliability factor and 95% confidence. This therefore provides documented endurance evidence for component evaluation. However, service life within an instrument still needs to be assessed against the intended liquids, electrical drive, contamination, and operating conditions.
Repeatable switching with a floating plunger
Parker describes a floating frictionless plunger and publishes a maximum cycling response of 10 ms. These features therefore provide a basis for repeatable switching in automated liquid circuits. However, actual liquid delivery remains dependent on the electrical command, pressure conditions, and surrounding flow path.
Match orifice size to pressure and flow
Parker’s published maximum pressure decreases as the orifice increases. Select against both required flow and pressure, rather than assuming every C15 configuration is rated to 10 bar.
| Orifice diameter | Maximum pressure | Published psi equivalent | Cv |
|---|---|---|---|
| 0.020 inch / 0.5 mm | 10 bar | 145 psi | 0.010 |
| 0.040 inch / 1.0 mm | 8 bar | 116 psi | 0.032 |
| 0.060 inch / 1.5 mm | 4 bar | 58 psi | 0.058 |
| 0.080 inch / 2.0 mm | 1.5 bar | 22 psi | 0.093 |
Use Parker’s water flow curves as a starting point, then test the intended liquid through the actual manifold. Viscosity, connecting passages, and downstream restrictions can change the delivered flow.
C15 liquid valve specifications
Use these family specifications to shortlist a valve, then confirm the exact part and manifold drawing before qualification.
| Specification | Published information |
|---|---|
| Valve function | Two port, normally closed solenoid cartridge valve. |
| Dimensions and weight | 15 mm diameter; 29 mm length; 22 g. |
| Wetted body materials | Series 300 and 400 stainless steel. |
| Seal options | FKM or EPDM, internal and external. |
| Standard coil voltage | 12 or 24 VDC, ±5%. |
| Typical rated power | 1.1 to 1.7 W, dependent on orifice and voltage. |
| Cycling response | 10 ms maximum. |
| Operating environment | 0 to 50°C. |
| Internal volume | 391 µL for the two way valve. |
| Recommended filtration | 10 µm. |
| Electrical connection | 80 mm flying leads, 24 AWG. |
| Installation | Manifold mounted cartridge seal interface with mechanical retention. |
| Published endurance rating | 500 million cycles; 0.90 reliability factor and 95% confidence. |
The above component specifications come from Parker’s C15 15mm Liquid Valve datasheet.
Learn more about Parker Miniature Valves and Components
Design the cartridge into your OEM assembly
Specify the entire wetted path
Choose seals against the actual process and cleaning liquids. For example, Parker’s abbreviated chart rates EPDM as excellent with methanol and acetonitrile, while FKM is not recommended for those media. The stainless steel components and manifold also need compatibility checks.
Include the published 391 µL valve volume alongside manifold and tubing volume when assessing liquid changeover and flushing. Prototype qualification should verify leakage, flow, switching performance, and removal of residual liquid under representative operating conditions.
Plan manifold sealing and mechanical retention
Follow the selected valve drawing for cavity dimensions and sealing surfaces. Parker specifies a maximum insertion force of 100 N and locates mechanical retention at the indent in the middle of the body. The securing clip should engage this feature.
Use compatible installation lubrication and confirm that insertion leaves both external O-rings correctly seated. Include access for the clip, electrical leads, and valve replacement in the assembly design to simplify maintenance within the instrument enclosure.
Match electrical drive to the switching sequence
For optional hit and hold operation, Parker specifies at least 100 ms at rated voltage before reducing to 50% of rated voltage. Its separate PWM guidance lists a minimum frequency of 1 kHz and a nominal holding duty of 50%.
Validate the holding requirement under the intended pressure, shock, vibration, and temperature conditions. The 100 ms initial drive requirement differs from the published 10 ms cycling response. Qualify reduced voltage and PWM control as separate drive methods.
Specify the configuration with confidence
the right configuration for procurement
Specify the exact Parker part number alongside the orifice, FKM or EPDM seals, coil voltage, electrical interface, and manifold arrangement. Include cavity, retaining clip, and driver requirements so the cartridge can be assessed as part of the assembly.
For development enquiries, provide sample quantities, process and cleaning liquids, inlet and outlet pressures, required flow, switching sequence, and qualification timetable. For production sourcing, include forecast volumes, delivery requirements, documentation, and agreed restrictions on substitutions.
Assess acquisition cost alongside manifold machining, driver hardware, assembly, qualification, and maintenance. Confirm current availability and lead times. The C15-MCS evaluation manifold, with its clip and screw, is ordered separately; Parker states that valves ship with O-rings installed.
Discuss your OEM cartridge valve requirements
AEG’s OEM Fluid Technology Division combines practical experience in medical and analytical fluid control with detailed knowledge of miniature liquid valve specification, material compatibility, and manifold integration. Our team can guide component selection and assembly development, considering orifice size, operating pressure, seal material, electrical drive, and cartridge retention.
Share your fluidic schematic, process and cleaning liquids, inlet and outlet pressures, required flow, switching sequence, electrical supply, available space, and expected quantities so we can assess your requirements and recommend suitable options.
Browse the Parker C15 liquid valve range, or contact Air Engineering Group on 01342 826488 or email info@aegroup.uk.com to discuss your medical, analytical, or life science liquid control application.
Find the right Parker valve for your OEM medical or life science application with guidance from Air Engineering Group. Whether you have a part number, a specification, or a fluid control challenge to solve, we can help you assess suitable options and move your project forward.
From selecting and specifying valves for prototype equipment to procurement and after sales support, AEG works with your engineering and purchasing teams to clarify requirements, address integration challenges, and match valves to your application.
Tell us about your application using the form below. Share the details you have, even if your requirements are still developing, and our team will help you identify the next step.